40
Teledyne API Model 465M O
3
Monitor User Manual
06160J DCN7969
7.2.5.
SETUP-VARS Menu
The VARS menu allows viewing and editing of various global setup variables that effect
how the 465M operates. These variables are stored in the instrument’s non-volatile
memory. Table 7-6 lists these variables and their function.
Table 7-6: VARS List
VAR
Meaning
Default
Value
Range
ALARM MODE
Controls the latching behavior of
the concentration alarms. See
Section 7.1.3 for additional
details.
LATCHING
LATCHING,
NON-
LATCHING
ALARM DELAY
Number of readings above alarm
limit required to trigger alarms
5
1 - 20
BC FILT SIZE
Boxcar filter length for
concentration filter. Higher
values will smooth out the
concentration readings but will
also increase the response time.
30
1 - 30
FLOW SWITCH
ENABLE
Enables/disables operation of
optional flow switch
OFF
ON, OFF
COMM ADDRESS
MODBUS address for COM
ports.
1
1 - 247
UNIT
Concentration unit-of-measure
setting.
PPM
PPM, WT%,
g/Nm
3
SCROLL DELAY
Adjustment for display scroll
delay. Higher values result in
slower scrolling speed.
20
6 - 100
AZERO ENABLE
Enables/Disables auto-zero
calibration
ON
ON/OFF
AZERO INTERVAL
Time interval for performing the
auto-zero calibration (minutes)
6
3 - 1440
GAS CARRIER
WEIGHT
Molecular weight of carrier gas
(ie. feed gas for ozone generator.)
This value is only used for the
calculation of wt% concentration
units. It is not used for PPM or
g/Nm3 calculations.
32
20.0 – 40.0
ANALOG RANGE
Full-scale concentration range.
Used to scale the analog output.
5000
5 – 50 g/Nm3
1000 – 9999 PPM
0.5 – 5.0 wt%
Содержание 465M
Страница 2: ...THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 12: ...10 Teledyne API Model 465M O3 Monitor User Manual 06160J DCN7969 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 24: ...22 Teledyne API Model 465M O3 Monitor User Manual 06160J DCN7969 Figure 4 6 NEMA Electrical I O Connections ...
Страница 30: ...28 Teledyne API Model 465M O3 Monitor User Manual 06160J DCN7969 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 32: ...30 Teledyne API Model 465M O3 Monitor User Manual 06160J DCN7969 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 44: ...42 Teledyne API Model 465M O3 Monitor User Manual 06160J DCN7969 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 70: ...68 Teledyne API Model 465M O3 Monitor User Manual 06160J DCN7969 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 72: ...70 Teledyne API Model 465M O3 Monitor User Manual 06160J DCN7969 THIS PAGE IS INTENTIONALLY LEFT BLANK ...