340CF ELSD
Section 3 Installation
3-4
4. Connect an inert gas supply (such as nitrogen) to the brass
fitting on the lower left rear panel of the ELSD, with
1
/
8
"
O.D. line.
Figure 3-4 Connecting gas supply to ELSD
5. Connect the external detector cable (Isco part
#250-0000-20) to the Rf200 (Figure 3-5).
6. Connect the other end of the cable to the ELSD, taking
care to observe polarity as indicated by the small tab on
the ground side of the connector.
Figure 3-5 Rf external detector cable connections
Ground
(CU)
Ground
Rf200 Connection
ELSD Connection