![Teledyne 311XL Series Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/teledyne/311xl-series/311xl-series_instruction-manual_1080430016.webp)
Accuracy at 25
o
C and constant pressure after thermal equilibrium has been reached.
Basic ac/- 2% of full scale for all ranges except 0-2 PPM range. Accuracy for 0-2 PPM
range (+/- 5% of full scale).
Accuracy over the temperature range 0-35C at constant pressure after thermal
equilibrium has been reached. Basic ac/- 5% of scale for all ranges except 0-2 PPM
range. Accuracy for 0-2PPM range (+/- 10% of full scale).
SENSITIVITY:
20ppb or better
REPRODUCIBILITY:
+/- 1% AT CONSTANT TEMPERATURE
RESPONSE:
90% in 90 sec. on the 0-2 PPM range with a flow rate of 2.5 SCFH.
SYSTEM OPPERATING TEMPERATURE:
32 – 98 DEG. F (0-35C)
SENSOR TYPE:
Micro fuel cell class B-2XL
SYSTEM POWER REQUIREMENTS:
100-115VAC 50/60 Hz (for battery charging only)
The system includes two sets of current limited NiCad batteries (internal).
WEIGHT:
6lb (2.71 kg).
-12-
RECOMMENDED SPARE PARTS LIST
311XL SERIES
QTY.
P/N
DESCRIPTION
1
F-39
MICRO-FUSE,
0.25
AMP.,
FAST-BLOW
(FOR
210-240
VAC)