If, on the other hand, the instrument is not to be used immediately after calibration, and a low 
ppm oxygen gas is being employed as a purge, allow the manifold to be purged overnight, and 
then disconnect both male fittings.  

ALWAYS DISCONNECT THE SOURCE FITTING FIRST, 

AND IMMEDIATELY THEREAFTER, THE VENT FITTING

 
 

3.3 

Positive Pressure Sampling:

 

When connecting the instrument to a positive 

pressure sample source, ALWAYS proceed as follows: 
 
 

 

1.  Before making ANY connections to the instrument, establish a sample line 

 

 

     flow rate of 0.1 to 10 liters/min.  Allow the sample to vent to atmosphere 

 

 

     long enough to purge the line free of air. 

 
 

 

2.  Install the vent fitting first and then the sample source fitting.  Be prepared 

 

 

     to make the connections in rapid order, so that atmospheric diffusion time 

 

 

     Through the vent fitting is held to a minimum. 

 
When disconnecting the instrument, reverse the procedure:  source fitting first, and then vent 
fitting. 
 
The objective of the connect-disconnect procedure is to obviate the possibility of pressurizing 
the manifold.  

IF A FLOWING SAMPLE WAS CONNECTED TO THE MANIFOLD 

WITHOUT THE VENT FITTING IN PLACE, THE PRESSURE IN THE MANIFOLD 
WOULD RISE AND BE EQUAL TO THE SAMPLE PRESSURE ALMOST 
IMMEDIATELY.

  In such a situation, depending on the magnitude of the sample pressure, 

leaks in the manifold might result. 
 
 

3.4 

Atmospheric Pressure Sampling:

 If the sample is at atmospheric pressure (or 

slightly negative), a sample pump will be required downstream from the analyzer.  The inlet side 
of the pump should also be equipped with a throttle valve, so that sample flow can be reduced to 
0.1 - 10 liters/min.  If pump loading is a consideration, the inlet side of the pump will have to 
include a bypass path that is open to the atmosphere through still another throttle valve.  The 
sample path and bypass path may then be balanced by manipulating the two valves, so that 
sample flow is within the prescribed limits without loading the pump. 
 
 
 

-6- 

UNDER NO CIRCUMSTANCES SHOULD THERE BY ANY RESTRICTIONS IN THE 
LINE BETWEEN THE SAMPLE POINT AND THE ANALYZER

 -- partial-pressure-

sensitive device, any oxygen readings taken under these conditions would be erroneous, and 
vacuums in excess of 1/3 atmosphere may damage the cell. 
 

4. MAINTENANCE

 

 

 

4.1 

Battery Power Supply Service.

  The Model 311D is for use ONLY when it is 

not connected to the AC power line.  TAI suggests that an overnight recharge be accomplished 

 

Содержание 311-D

Страница 1: ...MODEL 311 D RANGES 0 100 PPM 0 1000 PPM 1 AND 100 TELEDYNE ANALYTICAL INSTRUMENTS 16830 CHESTNUT STREET CITY OF INDUSTRY CALIF 91749 TELEPHONE 626 934 1500 FAX 626 961 2538 TWX 910 584 1887 TDYANYL CO...

Страница 2: ...MENT AND SERVICES 4 2 1 Sampling Equipment 4 2 2 Power Service 4 3 OPERATION 4 3 1 Introduction 4 3 2 Calibration 5 3 2 1 Calibration Procedure 5 3 3 Positive Pressure Sampling 6 3 4 Atmospheric Press...

Страница 3: ...to the oxygen concentration in the gas space immediately adjacent to the transducer sensing surface A two stage solid state amplifier that features operational amplifiers whose power consumption per u...

Страница 4: ...harging circuit A current limiting resistor 53 2 ohms is potted into the end of each battery This assures that under no circumstances can more than 80 milliamperes be switched or drawn from either bat...

Страница 5: ...fier maintains a constant ratio of 1 1 between the negative and positive power supplies The constant ratio keeps the offset voltage of the amplifier from changing as the battery voltage varies up and...

Страница 6: ...ce within the manifold and prior to shipment the manifold was purged with an inert gas to eliminate all but traces of oxygen from the internal sampling system The integral shutoff valves in the quick...

Страница 7: ...of sample flow is of no importance A pump is recommended on the plastic tube Pump the tube until the meter reading is stable CAUTION DO NOT SUCK ON THE TUBE WITH YOUR MOUTH THERE IS A POSSIBILITY THE...

Страница 8: ...E VENT FITTING IN PLACE THE PRESSURE IN THE MANIFOLD WOULD RISE AND BE EQUAL TO THE SAMPLE PRESSURE ALMOST IMMEDIATELY In such a situation depending on the magnitude of the sample pressure leaks in th...

Страница 9: ...moving parts in the instrument The Micro Fuel Cell is a sealed modular component that should be replaced when faulty 4 3 Cell Replacement The characteristics of the Micro Fuel Cell are similar to thos...

Страница 10: ...oved from its shipping package The customer should stock only one spare cell per instrument at a time Do not attempt to stockpile spare cells The Model 311D should not be used in applications where CO...

Страница 11: ...as possible 2 Place the vent line in water and disconnect the sample 3 Next disconnect the vent line and place the range switch on the X100 range 4 The unit should stay on the X100 range if there are...

Страница 12: ...ATION OF 100 PPM OR MORE FOR A CONCENTRATION OF LESS THAN 100 PPM THE SENSOR OUTPUT SHOULD BE WITHIN 10 IN SIXTY SECONDS PROVIDED THAT THE SENSOR IS NOT ALREADY SUPERSATURATED WITH A VERY HIGHLY CONCE...

Страница 13: ...EFFECT OF CO2 ON B 2C CELL LIFE 11...

Страница 14: ...ETER ASSEMBLY 1 L 79 LAMP NEON ASSEMBLY A MINIMUM CHARGE IS APPLICABLE TO SPARE PARTS ORDERS IMPORTANT Orders for replacement parts should include the part number if available model number serial numb...

Страница 15: ...311D B 26473 PICTORAL DIAGRAM C 28339 SCHEMATIC C 26606 WIRING DIAGRAM NOTE The MSDS on this material is available upon request through the Teledyne Environmental Health and Safety Coordinator Contact...

Страница 16: ......

Страница 17: ...Title Indsrsvr01 drawings MASTER 26000 C26606 14 dwg...

Страница 18: ...Title G MASTER 28000 C28339 8 DWG...

Отзывы: