Oxygen Analyzer
Part I: Control Unit
Part I: 4-1
Operation
4.1
Introduction
Once the analyzer has been installed, configure it for your process. To
do this you can:
•
Set system parameters—
•
Specify a password, if desired, requiring operator to log in.
•
Establish and start an automatic calibration cycle, if desired.
•
Calibrate the instrument.
•
Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.
•
Set alarm setpoints, and modes (latching, failsafe, etc).
Before configuration these default values are in effect:
PARAMETER
DEFAULT
LO Range
100 ppm
MED Range
1000 ppm
HI Range
10,000 ppm
Auto Ranging
ON
Alarm Relays
1000 ppm
(Defeated, HI, Not failsafe, Not latching)
Span
000008.00 ppm
(Auto, every 0 days at 0 hours)
Zero
(Auto, every 0 days at 0 hours).
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press
Enter
for access to all functions of the analyzer.
4.2
Using the Data Entry and Function
Содержание 3010TA
Страница 8: ...viii Model 3010 Model 3010 Model 3010 Model 3010 Model 3010T T T T TA A A A A...
Страница 12: ...iv Part I Model 3010TA Oxygen Analyzer...
Страница 18: ...1 6 Part I 1 Introduction Model 3010TA...
Страница 20: ...2 2 Part I 2 Operational Theory Model 3010TA Figure 2 1 Block Diagram of the 3010TA CU Electronics...
Страница 22: ...2 4 Part I 2 Operational Theory Model 3010TA...
Страница 66: ...1 4 Part II 1 Introduction 1 Introduction 1 Introduction 1 Introduction 1 Introduction Model 3010T...
Страница 80: ...3 Installation Model 3010T 3 8 Part II...
Страница 84: ...4 Operation Model 3010T 4 4 Part II...
Страница 92: ...5 Maintenance Model 3010T 5 8 Part II...