4  Operation

Model 3010

PAC

4-16:  Part I

 Teledyne Analytical Instruments

Use the 

∆∇

 arrow keys to enter the upper value of the range (all ranges

begin at 0 %). Repeat for each range you want to set. Press 

Enter

 to accept

the values and return to 

Analyze

 mode. (See note below.)

Note: The ranges must be increasing from low to high, for example,

if range 1 is set for 0–10 % and range 2 is set for 0–100 %,
range 3 cannot be set for 0–50 % since it is lower than range 2.

4.6.2 Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing the

analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, enter the range

function by pressing the 

Range

 button on the front panel.

Use the < > arrow keys to move the blinking over 

AUTO

.

Use the 

∆∇

 arrow keys to switch from 

AUTO

 to 

FX/LO

FX/MED, 

or

FX/HI

 to set the instrument on the desired fixed range (low, medium, or

high).

L—###    M—####
H—#####  Mode—FX/LO

or

L—###    M—####
H—#####  Mode—FX/MED

or

L—###    M—####
H—#####  Mode—FX/HI

Press 

Escape

 to re-enter the 

Analyze

 mode using the fixed range.

NOTE:When performing analysis on a fixed range, if the oxygen

concentration rises above the upper limit (or default value) as
established by the operator for that particular range, the
output saturates at 1 V dc (or 20 mA). However, the digital
readout and the RS-232 output continue to read the true value
of the oxygen concentration regardless of the analog output
range.

Содержание 3010PAC

Страница 1: ...OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PER SIST...

Страница 2: ...weight of any parts in so far as such alterations do not adversely affect our warranty ImportantNotice This instrument provides measurement readings to its user and serves as a tool by which valuable...

Страница 3: ...n Analyzer TeledyneAnalyticalInstruments Table of Contents Specific Model Information iv Preface v Part I Control Unit Model PAC Part I 1 1 Part II Intrinsic Safe Barriers Part II 1 1 Remote Probe App...

Страница 4: ...for which this manual was supplied are indicated by a check mark in the box InstrumentSerialNumber __________________________ 19 Rack Mount The 19 Relay Rack Mount units are available with either one...

Страница 5: ...te in a hazardous area Part I of this manual covers the Model 3010PAC General Purpose flush panel and or rack mount Control Unit only This Control Unit is for indoor use in a nonhazardous environment...

Страница 6: ...with bulkhead mount Control Unit Consists of 3010TB PB Control Unit PN D 66190A and a 3010T Analysis Unit PN D 65478 3010PB NECtypePercentOxygenAnalyzerwithbulkheadmount Control Unit Consists of 3010...

Страница 7: ...onics 8 bit CMOS microprocessor with 32 kB RAM and 128 kB ROM Three user definable output ranges allow best match to users process and equipment 0 1 through 0 100 Air calibration range for convenient...

Страница 8: ...ments of the Commonwealth of Europe CE for Radio Frequency Interference Electromagnetic Interference RFI EMI and Low Voltage Directive LVD The Analysis Unit is Intrinsically safe and CENELEC approved...

Страница 9: ...Part I i Part I Control Unit TeledyneAnalyticalInstruments OPERATING INSTRUCTIONS Models 3010PAC Oxygen Analyzer Part I Control Unit Flush Mount Part Number D 66192B...

Страница 10: ...1 Unpacking the Control Unit 3 1 3 2 Mounting the Control Unit 3 1 3 3 Electrical Connections 3 3 3 3 1 Primary Input Power 3 4 3 3 2 50 Pin Interface Connector 3 4 3 3 3 RS 232 Port 3 9 3 3 4 Remote...

Страница 11: ...4 2 Span Cal 4 11 4 4 2 1 Auto Mode Spanning 4 11 4 4 2 2 Manual Mode Spanning 4 12 4 5 The Alarms Function 4 13 4 6 The Range Function 4 15 4 6 1 Setting the Analog Output Ranges 4 15 4 6 2 Fixed Ra...

Страница 12: ...iv Part I Model 3010PAC Oxygen Analyzer TeledyneAnalyticalInstruments...

Страница 13: ...General Purpose flush panel and or rack mount Control Units The Analy sis Unit is covered in Part II of this manual The Control Unit is for indoor use in a nonhazardous environment only The Analysis U...

Страница 14: ...related tasks described in detail in chapter 4 Operation Span Span calibrate the analyzer Zero Zero calibrate the analyzer Alarms Set the alarm setpoints and attributes Range Set up the 3 user definab...

Страница 15: ...ating but there are no displays or outputs CAUTION The power cable must be unplugged to fully disconnect power from the instrument When chassis is exposed or when access door is open and power cable i...

Страница 16: ...oncentration alarms and 1 system alarm RS 232 Port Serial digital concentration signal output and control input Remote Probe Provides all electrical interconnect to the Analysis Unit or Remote Probe R...

Страница 17: ...sample system Network I O Serial digital communications for local network access For future expansion Not implemented at this printing Note If you require highly accurate Auto Cal timing use external...

Страница 18: ...1 6 Part I 1 Introduction Model 3010PAC Teledyne Analytical Instruments...

Страница 19: ...cations are illustrated in section 5 Maintenance Refer to Figure 2 1 Block Diagram of the 3010PAC CU Electronics In the presence of oxygen the sensor in the Analysis Unit generates a current A current...

Страница 20: ...2 2 Part I 2 Operational Theory Model 3010PAC Teledyne Analytical Instruments Figure 2 1 Block Diagram of the 3010PAC CU Electronics...

Страница 21: ...to the display alarms and communications port as well as to the optional gas control valves in the Analysis Unit The same digital information is also sent to a 12 bit digital to analog converter that...

Страница 22: ...2 4 Part I 2 Operational Theory Model 3010PAC Teledyne Analytical Instruments...

Страница 23: ...the system for operation Carefully unpack the Control Unit and inspect it for damage Immediately report any damage to the shipping agent 3 2 Mounting the Control Unit The Model 3010PAC Control Unit is...

Страница 24: ...t Figure 3 2 Single and Dual 19 Rack Mounts All operator controls are mounted on the control panel which is hinged on the left edge and doubles as a door to provide access to the internal components o...

Страница 25: ...3 Required Front Door Clearance 3 3 Electrical Connections Figure 3 4 shows the Control Unit rear panel Connections for power communications and both digital and analog signal outputs are described i...

Страница 26: ...trument s circuitry as long as the instrument is connected to the power source The switch on the front panel is for switching power on or off to the displays and outputs only Fuse Installation The fus...

Страница 27: ...oating 6 Range 4 20 mA floating 8 Range ID 0 1 V dc 23 Range ID 0 1 V dc negative ground 24 Range 0 1 V dc 7 Range 0 1 V dc negative ground Alarm Relays The three alarm circuit connectors are spring t...

Страница 28: ...44 Threshold Alarm 2 moving contact 43 Threshold Alarm 2 normally open contact 36 System Alarm normally closed contact 20 System Alarm moving contact 37 System Alarm normally open contact Digital Remo...

Страница 29: ...d and a zero or span command can be sent When the contact is CLOSED the analyzer is already calibrating It will ignore your request to calibrate and it will not remember that request Once a zero or sp...

Страница 30: ...n connections Table 3 4 Range ID Relay Connections Pin Function 21 Range 1 ID Contact 38 Range 1 ID Contact 22 Range 2 ID Contact 39 Range 2 ID Contact 19 Range 3 ID Contact 18 Range 3 ID Contact 34 R...

Страница 31: ...e obtainable voltage depending on the load impedance applied See Figure 3 6 Figure 3 6 FET Series Resistance 3 3 3 RS 232 Port The digital signal output is a standard full duplex RS 232 serial communi...

Страница 32: ...010PAC implementation Table 3 6 Required RS 232 Options Parameter Setting Baud 2400 Byte 8 bits Parity none Stop Bits 1 Message Interval 2 seconds 3 3 4 Remote Probe Connection The Models 3010PAC are...

Страница 33: ...oth installed and interconnected and the system gas and electrical connections are complete the system is ready to test Before plugging either of the units into their respective power sources Check th...

Страница 34: ...3 Installation Model 3010PAC 3 12 Part I Teledyne Analytical Instruments...

Страница 35: ...ysis ranges Then choose autoranging or select a fixed range of analysis as required Set alarm setpoints and modes latching failsafe etc Before configuration these default values are in effect PARAMETE...

Страница 36: ...operating mode The analyzer monitors the oxygen content of the sample displays the concentration of oxygen and warns of any alarm conditions System The system function consists of six subfunctions tha...

Страница 37: ...shbutton names are printed in Oblique type 4 3 The System Function The subfunctions of the System function are described below Specific procedures for their use follow the descriptions Auto Cal Used t...

Страница 38: ...ique password is assigned to defeat the security the password must be changed back to TBEAI Logout Logging out prevents an unauthorized tampering with analyzersettings More Select and enter More to ge...

Страница 39: ...o set OFF to ON if a zero interval is entered A Span Every or Zero Every screen appears Span Every 0 d Start 0 h from now Use arrows to set an interval value then use arrows to move to the start time...

Страница 40: ...roll the blinking over to PSWD and press Enter to select the password function Either the default TBEAI password or AAAAA place holders for an existing password will appear on screen depending on whet...

Страница 41: ...default TBEAI or the previously assigned password or press Escape to keep the existing pass word and move on If you chose Enter to change the password the password assignment screenappears T B E A I E...

Страница 42: ...0 100 If an alarm is tripped the second line will change to show which alarm itis 0 0 Anlz A L 1 NOTE If you previously logged off the system you will now be required to re enter the password to gain...

Страница 43: ...n by the operator at will To initiate a self diagnostic test during operation Press the System button to start the System function Contrast Auto Cal PSWD Logout More Use the arrow keys to blink More t...

Страница 44: ...ll scale oxygen analysis immediately following calibrationinair Connect the calibration gases to the analyzer according to the instruc tions given in Section 3 4 1 Gas Connections observing all the pr...

Страница 45: ...n also be performed in manual mode where the operator deter mines when the span concentration reading is acceptable and manually exits thefunction 4 4 2 1 AutoModeSpanning Press Span to enter the span...

Страница 46: ...tration of the span gas you are using 20 90 if you are using air The arrows choose the digit and the arrows choose the value of the digit Press Enter to enter the span value into the system and begin...

Страница 47: ...ints 1 Which if any of the alarms are to be high alarms and which if any are to be low alarms Setting an alarm as HIGH triggers the alarm when the oxygen concentration rises above the setpoint Setting...

Страница 48: ...ng over to AL 1 using the arrow keys Then press Enter to move to the next screen AL 1 10 HI Dft N Fs N Ltch N Five parameters can be changed on this screen Value of the alarm setpoint AL 1 oxygen Out...

Страница 49: ...low 85 of full scale of the next lower range the instrument will switch to that range A corresponding shift in the DC percent of range output and in the range ID outputs will be noticed The autorangin...

Страница 50: ...nge To switch from autoranging to fixed range analysis enter the range function by pressing the Range button on the front panel Use the arrow keys to move the blinking over AUTO Use the arrow keys to...

Страница 51: ...lternatively you can press the Analyze button at any time to return to analyzing your sample 4 8 Signal Output The standard Model 3010PAC Oxygen Analyzer are equipped with two 0 1 V dc analog output t...

Страница 52: ...sary to know what range the instrument is currently on especially when the analyzer is in the autoranging mode To provide an indication of the range a second pair of analog output terminals are used T...

Страница 53: ...cing Micro Fuel cells and replacing fuses in the Analysis Unit are covered in Part II Chapter 5 For recalibration see Part I section 4 4 Calibration WARNING SEE WARNINGS ON THE TITLE PAGE OF THIS MANU...

Страница 54: ...Test 1 Press the System button to enter the system mode 2 Use the arrow keys to move to More and press Enter 3 Use the arrow keys to move to Self Test and press Enter The following failure codes appl...

Страница 55: ...ED The 3010PAC Control Units contain the following major components Power Supply Motherboard with Microprocessor RS 232 chip and Preamplifier PCB Front Panel Display Board and Displays 5 digit LED met...

Страница 56: ...me of cleaning disconnect the instru ment from the power source Close and latch the front panel access door Clean outside surfaces with a soft cloth dampened slightly with plain clean water Do not use...

Страница 57: ...PartII i Part II Analysis Unit TeledyneAnalyticalInstruments OPERATING INSTRUCTIONS Model 3010PAC Oxygen Analyzer Part II Intrinsic Safe Barriers and Remote Probe CENELEC Type Part Number C 66426...

Страница 58: ...Pressure 2 4 2 2 5 Calibration Characteristics 2 4 2 2 6 Micro Fuel Cell Class 2 5 2 2 6 1 Class A 3 Cell 2 5 2 2 6 2 Class A 5 Cell 2 5 2 2 6 3 Class B 1 Cell 2 6 2 2 6 4 Class B 3 Cell 2 6 2 2 6 5...

Страница 59: ...ically and are intended for applica tions where remote control of the sample analysis is desirable Safety barrier box isolate the BASEEFA approved probes from the control unit The intrinsically safe s...

Страница 60: ...barriers P N B366 and P N B367 and barrier box P N E324 as shown in drawing C66191 1 4 Cell Housing Probe The cell housing probe provides an enclosure for the B1 sensor and temperature compensation th...

Страница 61: ...sed in the remote Control Unit covered in Part I of this manual 2 2 Micro Fuel Cell Sensor 2 2 1 Principles of Operation The oxygen sensors used in the Model 3010PAC series are Micro Fuel Cells design...

Страница 62: ...idently in practically any environment or sample stream The cell is effectively sealed although one end is permeable to oxygen in the sample gas The other end of the cell is a contact plate consisting...

Страница 63: ...rear membrane beneath the anode is filled with electro lyte Cathode and anode are submerged in this common pool They each have a conductor connecting them to one of the external contact rings on the c...

Страница 64: ...rease The electron transfer and therefore the external current will increase even though the oxygen concentration of the sample has not changed It is therefore important that the sample pressure at th...

Страница 65: ...pical use and electri cal specifications 2 2 6 1 Class A 3 Cell The class A 3 cell is for use in applications where it is exposed con tinuously to carbon dioxide concentrations between 1 and 100 in th...

Страница 66: ...ptable in order to have a longer cell life Nominal output in air is 0 30 mA and 90 response time is 13 s Expected life in air is 12 months 2 2 6 5 Class C 3 Cell The class B 1 cell is for use in appli...

Страница 67: ...that is sheltered from the elements In areas where the ambient temperature is expected to drop below 0 C auxiliary heating must be provided Install the probe with a suitable sample system in an area...

Страница 68: ...the instrument has been idle for too long When the micro Fuel Cell needs to be installed or replaced follow the procedures in chapter 5 Maintenance for removing and installing cells 3 4 Testing the Sy...

Страница 69: ...ic safety barrier strips one each P N B366 single channel B367 dual channel are installed between the cell and the control unit The barrier strips are housed in an approved bulkhead mountable barrier...

Страница 70: ...0 1 V 13 Network 15 Zero Solenoid Return 16 Span Solenoid Return 17 Span Solenoid Hot 18 Range 3 Contact 19 Range 3 Contact 20 Alarm 3 C Contact 21 Range 1 Contact 22 Range 2 Contact 23 Range ID 0 1...

Страница 71: ...O Contact 48 Exhaust Solenoid Return 49 Zero Solenoid 50 Sample Solenoid Return 3 5 Remote Probe Connection The Models 3010PAC are split architecture dual chassis instru ments which have a Remote Prob...

Страница 72: ...3 Installation Model 3010PAC 3 6 Part II Teledyne Analytical Instruments...

Страница 73: ...ON THE TITLE PAGE OF THIS MANUAL 4 2 When to Replace a Cell If the 3010P analysis readings begin to drift downward uncharacteristi cally try recalibration If recalibration raises the readings for a sh...

Страница 74: ...concentration of your span gas If the quotient is less than the Index value for the cell class you are using replace the cell Table 5 1 Cell Indices Cell Class Index A 3 1 818 A 5 1 818 B 1 4 545 B 3...

Страница 75: ...ou Note Evidence of damage due to tampering or mishandling will render the cell warranty null and void 4 4 System Self Diagnostic Test 1 Press the System button to enter the system mode 2 Use the arro...

Страница 76: ...4 Maintenance Model 3010PAC 4 4 Part II Teledyne Analytical Instruments...

Страница 77: ...endix A 1 Oxygen Analyzers Appendix TeledyneAnalyticalInstruments OPERATING INSTRUCTIONS Model 3010PAC Oxygen Analyzers Appendix Flush Mount Control Unit PN D66192B CENELEC Type Analyssis Unit PN B399...

Страница 78: ...2 Appendix Appendix Models 3010PAC TeledyneAnalyticalInstruments Contents A 1 Model 3010PAC Specifications A 3 A 2 Recommended 2 Year Spare Parts List A 5 A 3 Drawing List A 6 A 5 The Zero Functions A...

Страница 79: ...Threeuserdefinableranges0 1 to0 100 Air calibration range 0 25 Autoranging with range ID output SampleSystem Positivepressureservice Vacuumservice optional Auto Cal Auto Zero Electrically operated va...

Страница 80: ...to meet but not yet certified to EN 50081 1 EN 50082 2 Accuracy 2 of full scale at constant temperature 5 of full scale over operating temperature range on factory default analysis ranges once therma...

Страница 81: ...66 Safety Barrier 1 Channel 1 B367 Safety Barrier 2 Channel _____________________ Order one type only A B or C as appropriate Order one type only US or European as appropriate Check Specific Model Inf...

Страница 82: ...ntrol Unit 3010PAC D 66192C Final Assembly Outline Drawing Control Unit 3010PAC MA D66191 Sensor Block Interconnection Diagram C66426 Safety Barrier Housing Assembly B39923C Final Assembly Top 4 Probe...

Страница 83: ...appears blinking on the display Zero Settling AUTO ENT To Begin Press Enter to begin zeroing Zero Slope ppm s The beginning zero level is shown in the upper left corner of the dis play As the zero rea...

Страница 84: ...ys the rate of change as Slope a value in parts per million per second ppm s Zero Slope ppm s NOTE It takes several seconds for the true Slope value to display Wait about 10 seconds Then wait until Sl...

Страница 85: ...2 Environmental Health and Safety Extension 230 Date Prepared 04 26 95 Section I Product Identification Section III Physical Chemical Characteristics Section II Hazardous Ingredients Composition Mater...

Страница 86: ...utes of Entry Inhalation Highly unlikely Ingestion May be fatal if swallowed Skin The electrolyte potassium hydroxide is corrosive skin contact may cause irritation or chemical burns Eyes The electrol...

Страница 87: ...inutes and remove all contaminated clothing Get medical attention immedi ately If ingested give large amounts of water and DO NOT INDUCE VOMITING Obtain medical attention immedi ately If inhaled remov...

Страница 88: ...mation RQ Hazardous Waste Solid N O S Lead Class 9 NA3077 PG III Follow all Federal State and Local regulations Section X References Material Safety Data Sheets from J T Baker Chemical Aldrich Malinck...