Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
3.4 Automatic Zero/Span Check
Automatic zero/span checking (Z/S check) must be enabled in the setup mode. The Teledyne API
Model 300M Carbon Monoxide Analyzer with Zero/Span Valves option offers capability to check
the zero and span point automatically on a timed basis, or through remote RS-232 operation (see
Section 3.7).
Under the
SETUP-ACAL
menu, there are three separate auto-sequences called SEQ1, SEQ2, and
SEQ3. Under each SEQ, there are four setup parameters that affect zero/span checking: the mode,
the starting date of the check, the time of day for check, the number of delay days, time delay,
the duration of the check, the range to check and whether to calibrate as well as check the range.
These are described individually below. Use the
PREV
and
NEXT
buttons to scroll through the
three sequences. The mode for each sequence is displayed. To change the mode for any of the
sequences, scroll to the desired sequence and press the
MODE
button. Use the
PREV
and
NEXT
buttons to select one of the modes shown below and press
ENTR
.
Mode:
1. DISABLED (Sequence is disabled)
2. ZERO
3. SPAN
4. ZERO-SPAN
To change the setup parameters for a sequence, press the
SET
button. Pressing the
<SET
and
SET>
buttons allows you to scroll through the setup parameters and edit them by pressing the
EDIT
button. The function of each setup parameter is described below:
Timer Enabled:
When set to
ON
, the sequence will be executed based on the internal timer, as
specified in the following parameters. When set to
OFF
the sequence will be executed only upon
an external (RS-232) command.
Starting Date:
The starting date for the sequence is entered in the format of MM/DD/YY, where
MM is the month, DD is the date, and YY is the year. Enter starting date and press
ENTR
or
EXIT
to leave the date unchanged.
Starting Time:
To set the time of day for the sequence, enter in the format HH:MM, where HH
is the hour in 24-hour format (i.e. hours range from 00 to 23) and MM is the minute (00 - 59).
Enter the time of day for calibration check and then press
ENTR
to accept the new time or
EXIT
to leave the time unchanged.
3-5
Содержание 300M
Страница 12: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A Figure 1 1 Pneumatic Diagram 1 4 ...
Страница 17: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A Figure 1 2 Carbon Monoxide Analyzer 1 9 ...
Страница 26: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 1 18 ...
Страница 36: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 2 10 ...
Страница 56: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 4 10 ...
Страница 64: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 5 8 ...
Страница 66: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 6 2 ...
Страница 86: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 8 8 ...
Страница 94: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 9 8 ...
Страница 116: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 11 4 ...