Teledyne 202 Series Скачать руководство пользователя страница 21

200/202 Series Flowmeters and Controllers Instruction Manual

 

EAR99 Technology Subject to Restrictions Contained on the Cover Page

 

Page 21 of 27 

3. 

Maintenance 

This section contains service and calibration information.  Some portions of the instrument are delicate, so 
please use extreme care when servicing the flow controller. 

NOTICE:

  It may prove more prudent in the long run to delegate the maintenance of Hastings flow 

instruments to Hastings personnel with the appropriate equipment and training.  Improper adjustments 
to the flow instruments can result in inaccurate data, poor instrument function, and may lead to 
sending the unit back for repair. 

 

Authorized Maintenance 

With proper care in installation and use, the flow controller will require little or no maintenance. If 
maintenance does become necessary, most of the instrument can be cleaned or repaired in the field. Some 
procedures may require recalibration. Do not attempt these procedures unless facilities are available. Entry 
into the sensor or tampering with the printed circuit board will void the warranty. Do not perform repairs on 
these assemblies while unit is still under warranty. 

 

Troubleshooting 

This section includes some of the more common issues that may arise when using a Hastings flow meter or 
controller.  Symptoms, causes, and recommended actions are grouped with each problem. 
 

SYMPTOM:

 Flow remains when 0.00 VDC is commanded or even when Override CLOSE function is enabled. 

CAUSE:

 Orifice out of adjustment or faulty op-amp 

ACTION:

 Check the valve voltage between pins 2 & 12 on the 15-pin D-connector for 15 Volt units and 

Pins 5 & 8 on the DE-9 connector for 24 Volts. If the voltage is less than 3.00 VDC, then turn the orifice 
slowly clockwise until flow stops. 

 

NOTICE:

  adjusting the orifice too quickly or two far can damage the valve seat and cause leak paths 

that will not allow the controller to operate properly. Please adjust the orifice gently and with caution. 
 

SYMPTOM:

 Output of unit is proportional to flow but extremely small and not correctable by span pot. 

CAUSE:

 Sensor is not being heated. 

ACTION:

 Unplug the sensor connector (J4 for 15 V boards, J5 for 24 V boards) and check the following 

resistances: The resistance between pins 2 & 3 of the sensor should be about 

2.475 kΩ ±75 Ω

. The resistance 

between pins 1 & 4 should be somewhere around 2 to 3 

Ω

. The resistance between pin 2 or 3 and the base of 

the sensor should be essentially infinite. If not, the heater wire is contacting the sensor base and the sensor 
must be replaced. If the sensor passes the previous three tests, check the voltage output on pins 2 & 3 of the 
sensor connector on the PCB. If it does not read approximately 22 to 23 VDC then replace the sensor heater 
regulator (U1 on 15 V board, U2 on 24 Volt board). 

 

SYMPTOM:

 Sensor has proper resistance readings, but little or no output with flow. 

CAUSE:

 Plugged sensor. 

ACTION:

 Shut off gas supply and power supply. Remove cover and PC board from unit. Remove 

sensor assembly and examine. If sensor has evidence of plugging, clean or replace as applicable 

 

SYMPTOM:

 Flow controller oscillates. 

CAUSE:

 Flow controller not adjusted for the dynamics of the flow system. 

ACTION:

 Check upstream and downstream pressures. The gas supply regulator should not have 

excessive lockup when flow shuts off. Also ensure that there is not a large drop in pressure between the 
regulator and the instrument due to line resistance. Oscillations can also be caused if a large flow 
restriction is pneumatically close to the downstream end of the flow controller. The differential pressure 
across the unit must be between 10-50 psig. 

 

SYMPTOM:

 Little or no flow, even with Valve Override OPEN enabled. 

Содержание 202 Series

Страница 1: ...ION MANUAL www teledyne hi com I S O 9 0 0 1 C E R T I F I E D Use and Disclosure of Data Information contained herein is classified as EAR99 under the U S Export Administration Regulations Export reexport or diversion contrary to U S law is prohibited ...

Страница 2: ... Number 140 102009A October 2009 Revision L Document Number 140 082010 August 2010 Revision M Document Number 140 072011 July 2011 Revision N Document Number 140 072013 July 2013 Revision P Document Number 140 032021 March 2021 Visit www teledyne hi com for WEEE disposal guidance Description of Symbols and Messages used in this manual WARNING indicates a hazardous situation which if not avoided co...

Страница 3: ... 4 3 Upstream Differential Pressure 12 2 4 4 Other Mounting Fitting Options Notes 12 2 4 5 Laminar Flow Element LFE Configuration 13 ELECTRICAL CONNECTIONS 14 2 5 1 Bipolar 15 VDC Power Supply to Bipolar Connections 14 2 5 2 Unipolar 24 VDC Power Supply to Unipolar Connections 15 2 5 3 General Connection Notes 15 OPERATION 16 2 6 1 Operating Conditions 16 2 6 2 Zero Check 16 2 6 3 High Pressure Hi...

Страница 4: ...ROL 20 3 MAINTENANCE 21 AUTHORIZED MAINTENANCE 21 TROUBLESHOOTING 21 ADJUSTMENTS 22 3 3 1 Calibration Procedure controller 22 3 3 2 Miscellaneous Adjustments 23 ENDCAP REMOVAL 23 PRINTED CIRCUIT BOARD REPLACEMENT 23 SENSOR REPLACEMENT 23 ORIFICE CHANGES 23 3 7 1 HFC 202 Orifice 24 4 WARRANTY AND REPAIR 25 WARRANTY REPAIR POLICY 25 RETURN POLICY 25 5 DRAWINGS 26 ...

Страница 5: ...m the use of moderate overpressures 500 psi 3 45MPa or overflows Instruments are normally calibrated with the appropriate standard calibration gas nitrogen then a correction factor is used to adjust the output for the intended gas NOTICE The instruments described in this manual may be available with multiple pinouts Ensure that all electrical connections are correct NOTICE The instruments describe...

Страница 6: ... and span points MODULAR SENSOR WITH LOCKING CONNECTOR The HFM 200 HFC 202 Series incorporates a removable replaceable sensor module with a reliable sensor to board connection Field repairs to units can be achieved with a minimum of production line downtime with confidence in the sensor to PCB connection SETTLING TIME Filtering in the meter circuitry and more modern valve circuitry have improved t...

Страница 7: ...6 VDC 20 mA 0 88 watts Power 15V meter 4 20 mA 14 16 VDC 55 mA 14 16 VDC 20 mA 1 2 watts Power 24V meter 0 5 VDC 11 36 VDC 1 08 watts Power 24V meter 4 20 mA 11 36 VDC 1 80 watts NOTICE Use of an underpowered or under voltage supply could result in equipment damage Output Flow Signal 0 5 00 VDC linear or 4 20 mA Wetted Material2 316 SS Viton 82 18 Au Ni Braze Trace Silver Solder Connector 15 pin s...

Страница 8: ...1 mA 14 16 VDC 175 mA 3 14 watts Power 15V controller 4 20 mA 14 16 VDC 40 mA 14 16 VDC 175 mA 3 44 watts Power 24V controller 0 5 VDC 12 36 VDC 3 64 watts Power 24V controller 4 20 mA 13 36 VDC 3 96 watts NOTICE Use of an underpowered or under voltage supply could result in equipment damage Output Flow Signal CMD Input 0 5 00 VDC linear or 4 20 mA Wetted Material2 316 302 SS Ni Viton Kalrez 82 18...

Страница 9: ...E 9 24 Volt The current loop is sourced internally from the 15 VDC supply and must be returned to the signal supply common after pin 5 15 VDC or pin 8 24 VDC passing through the monitoring circuitry to complete the loop The load must be between 0 and 600 Ohm The 4 20 mA I O option can accept a current input for flow controllers The 0 5 VDC command signal on pin 14 pin 3 for 24 Volts can be replace...

Страница 10: ...f switch not be blocked Throughout this manual when reference is made to a power supply it is assumed the customer is using a Hastings power supply Designs and models offered as Hastings Power supplies change as new technology becomes available For knowledge of the most up to date Hastings power supplies please consult either Hastings sales staff or our website at http www teledyne hi com WARNING ...

Страница 11: ...equirements See specifications in sections 1 2 and 1 3 The HFM 200 HFC 202 Series supply voltages should be reasonably regulated as power supply ripple may propagate to the output Surge suppressors are recommended to prevent power spikes reaching the instrument The Hastings power supplies in section 1 5 1 satisfy these requirements Hastings power supplies are limited to no more than 150VA max and ...

Страница 12: ...diameter of the sensor tube 0 0125 0 31 mm so the instrument requires adequate filtering of the gas supply to prevent blockage or clogging of the tube 2 4 3 Upstream Differential Pressure The pressure regulator and the plumbing upstream must be of sufficient size to minimize changes in the upstream pressure In general high capacity regulators and large internal diameter plumbing help to make the s...

Страница 13: ... after the sensor outlet tap To properly apply the rule of thumb we also need to know the distance between the LFE edges and the sensor taps LFE ID Edge to Sensor Tap LFE length 3 inches 1 5 inches 8 inches 3 inches 2 5 inches 10 inches Example Suppose a 4 inch LFE is selected The length of 4 inch inner diameter tubing or pipe upstream of the sensor inlet tap will be a minimum of 20 inches The min...

Страница 14: ...strument and void the warranty The figure below shows the schematic layout for connecting the instrument to an appropriate 15 Volt power supply A pinout is also given to show the signals available to the user and the signals given and used by the power supply Not all signals from the PCB are used by Hastings supplies The power supply used must be meet the power requirements delineated in the speci...

Страница 15: ...f the DE 9 24 Volt is the command input This should be a 0 5 VDC or 4 20 mA signal and must be free of spikes or other electrical noise as these will generate false flow commands that the controller would attempt to adjust for Pin 15 of the DA 15 15 Volt and pin 1 of the DE 9 24 Volt is a regulated 5 VDC output 0 005 V reference which is occasionally used as a full scale command If a valve overrid...

Страница 16: ...t already energized Allow for a 1 hour warm up Stop all flow through the instrument and wait 2 minutes Caution Do not assume that all metering valves completely shut off the flow Even a slight leakage will cause an indication on the meter and an apparent zero shift For the standard 0 5 VDC output adjust the zero potentiometer located on the inlet side of the flowmeter until the meter indicates zer...

Страница 17: ...ow rate of 100 SLM Controller B is a 5 V controller with a full scale flow rate of 10 SLM What gain factor on the output of controller A will result in controller B flowing 5 of controller A Use the formulas above 𝐺𝑎𝑖𝑛 𝐹𝑎𝑐𝑡𝑜𝑟 𝑜𝑓 𝐴 0 05 100 5 10 5 0 5 Cut the output of controller A in half and route the signal to the setpoint of controller B to flow 5 of controller A Operation with External Devices...

Страница 18: ...2 WARNING Care must be taken to avoid any high voltages that may be present when dealing with power supplies To successfully operate a flow controller a 0 5 VDC command signal proportional to the intended flow 0 Volts zero flow 5 Volts 100 of rated flow must be applied to pin 14 of the DA 15 15 Volt Pin 3 of the DE 9 24 Volt connector A 0 5 VDC signal proportional to the flow rate through the inst...

Страница 19: ...ft side of the PCB between the span potentiometer R7 and the large D connector J1 In the normal operating mode the jumper covers the left two pins To select an external sensor input move the jumper to the right two pins This swaps the flow input to the controller circuit from the flow meter output to pin 13 of the DA 15 connector 2 7 3 2 24 Volt Units For the 24 Volt units J8 is a six pin jumper b...

Страница 20: ... 100 and observe the flow through the controller If the overshoot is tolerable make no further changes Else remove the jumper and to prevent losing the unused jumper place it over only one pin of J15 one of the pins of J8 for 24 V boards Range Changing The range of the flow controller can be changed in the field if recalibration facilities are available The flow controller may require a different ...

Страница 21: ...low stops NOTICE adjusting the orifice too quickly or two far can damage the valve seat and cause leak paths that will not allow the controller to operate properly Please adjust the orifice gently and with caution SYMPTOM Output of unit is proportional to flow but extremely small and not correctable by span pot CAUSE Sensor is not being heated ACTION Unplug the sensor connector J4 for 15 V boards ...

Страница 22: ...f a calibration reference 1 Connect power to the instrument put in AUTO state and allow 10 flow for 30 minutes as a warm up 2 After the warm up period set the flow to zero and wait for 2 minutes Check to make sure there is zero flow 3 Adjust the ZERO potentiometer bottom hole on inlet side so that the flow instrument outputs 0 000 VDC 6 5 on 15V 2 8 on 24V 4 Override the valve closed or give a 0 f...

Страница 23: ... two jackscrews next to the D connector and the two screws on the sides of the cover Lift off the cover and unplug the four wire sensor plug and the two wire valve plug noting their orientation prior to removal Remove the screw that holds the PC board to the sensor Troubleshoot or replace as applicable Installation is the reverse of the above procedure Recalibrate if any components were changed or...

Страница 24: ...ght out See Figure in section 3 7 Prior to reinstallation of the orifice inspect the two O rings mounted on it for damage Replace if cut or gouged Lubricate the O rings slightly with a silicone based grease and the threads with anti galling compound Push the orifice into its hole and screw it in until it is flush with the instrument base Apply pressure to the inlet side of the instrument Enable th...

Страница 25: ...ty or non warranty repair must have been issued a Return Material Authorization RMA form RMA Forms may be obtained from the Information section of the Hastings Instruments website http www teledyne hi com inforeq htm WARNING Contaminated parts can be detrimental to health and environment Ensure instruments are free of hazardous contamination prior to shipment Company contact information TELEDYNE H...

Страница 26: ...200 202 Series Flowmeters and Controllers Instruction Manual EAR99 Technology Subject to Restrictions Contained on the Cover Page Page 26 of 27 5 Drawings ...

Страница 27: ...200 202 Series Flowmeters and Controllers Instruction Manual EAR99 Technology Subject to Restrictions Contained on the Cover Page Page 27 of 27 ...

Отзывы: