Model 200E Instruction Manual
Operating Instructions
The status outputs are accessed through a 12 pin connector on the analyzer’s rear panel
labeled STATUS. The function of each pin is defined in Table 6-14.
COMMON
EMIT
T
E
RS
STATUS
1 2 3 4 5 6 7 8 D
+
SY
ST
E
M
O
K
HI
GH
RA
NGE
CONC
V
A
LI
D
ZE
R
O
CA
L
SP
A
N
CAL
D
IAG
NO
ST
IC MO
DE
LO
W SPAN
GR
OU
N
D
Figure 6-4: Status Output Connector
Table 6-14: Status Output Pin Assignments
Connector
Pin
Status Condition
(ON=conducting)
1
SYSTEM OK
ON if no faults are present.
2
CONC VALID
ON if concentration measurement is valid, OFF when invalid.
3
HIGH RANGE
ON if unit is in high range of any AUTO range mode.
4
ZERO CAL
ON whenever the instrument is in ZERO calibration mode.
5
SPAN CAL
ON whenever the instrument is in SPAN calibration mode.
6
DIAG MODE
ON whenever the instrument is in DIAGNOSTIC mode.
7
LOW RANGE
ON if unit is in low range of any AUTO range mode.
8
Unused.
D EMITTER
BUS
The emitters of the transistors on pins 1-8 are bussed
together. For most applications, this pin should be connected
to the circuit ground of the receiving device.
+ DC
POWER + 5 VDC, 30 mA maximum (combined rating with Control
Inputs).
DIGITAL
GROUND
The ground from the analyzer’s internal, 5/±15 VDC power
supply.
6.8.2. Control Inputs
Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes are
provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.
These are opto-isolated, digital inputs that are activated when a 5 VDC signal from the “U”
pin is connected to the respective input pin.
044100102 Rev A
73
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...