Maintenance
2000 XTC
Maintenance
4.1 Routine Maintenance
Aside from normal cleaning and checking for leaks at the gas
connections, routine maintenance is limited to recalibration. For
calibration procedures, refer to Section 3.6. On occasion, it may be
necessary to replace a blown fuse. This is covered in section 4.2.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
The overall design of the instrument is intended to facilitate servicing
and troubleshooting, should that ever be necessary. The sensor itself is a
solid-state unit with no serviceable parts. The cell is enclosed in an
insulated compartment that is readily accessible after removing the
transmitter cover. There are three (3) PC boards handling the electronics:
one in the transmitter module and two in the galvanic isolator module. Both
boards are easily accessible after removing the respective front covers.
4.2 Troubleshooting
The following table offers guidelines for diagnosing and correcting
common problems associated with the 2000 XTC Thermal Conductivity
Transmitter.
Table 4-1 Troubleshooting
Symptom Cause
Remedy
The unit will not
power up.
Power not connected
or a wire has pulled
loose.
A) Make sure 24VDC power
is connected and turned
on.
B) Check input power wiring
at rear of galvanic isolator
module.
Teledyne Analytical Instruments
22