4 Operation
Model 1220
4-6
Teledyne Analytical Instruments
Configuration Display Panel Fields
“S1” highlighted indicates that Sensor 1 options may be set.
“S2” highlighted indicates that Sensor 2 options may be set.
This option is used to zero (“0”) calibrate the Channel Module.
This option is used to span calibrate the Channel Module.
This option enables a single person to perform both a span and
zero calibration.
“H” highlighted indicates that the High Alarm value can be set
by using the Up/Down Arrow Keys.
“C” highlighted indicates that the Caution Alarm value can be
set by using the Up/Down Arrow Keys.
When the instrument goes into an alarm condition, the alarm
relay switches. If “LCH” (latched) was selected, the alarm relay
remains switched (latched) even after the alarm condition has
been cleared. Using this configuration, you must press the
RESET switch to unlatch (unlock) the relay.
When the instrument goes into an alarm condition, the alarm
relay switches. If “NLCH” (non-latched) was selected, the
alarm relay switches (unlatches) when the alarm condition is
cleared.
NOTE:
When one of these modes
are selected, the field light
flashes indicating which
mode the monitor is on.