![Teledyne SP Devices ADQ8-8C Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/teledyne-sp-devices/adq8-8c/adq8-8c_manual_1081092021.webp)
17-2000 C
2020-09-17
21(50)
ADQ8-8C Manual
17-2000 C 2020-09-17
21(50)
2.
DisArmTrigger
3.
SetTriggerMode
(“software trigger”)
4.
ArmTrigger
5.
SWTrig
6. Read data and analyze the situation
4.8 External Trigger Inputs
An external trigger is a dedicated signal on a dedicated input to the ADQ. There are several inputs for
external trigger,
.
4.8.1
External trigger TRIG front panel connector
The block diagram for the TRIG is shown in
and related parameters are listed in
. The
user can control the external trigger function for adapting it to the system in the following ways:
•
The input impedance can be set in 50
(default) or high impedance 500
mode, see
•
Configure the threshold level.
•
Set the trigger edge to rising or falling to adjust to the polarity of the trigger signal.
CONNECTOR
DESCRIPTION
TIME
RESOLUTION
TOTAL
JITTER
IMPEDANCE
TRIG
LEVEL
REF
TRIG
External trigger on
front panel.
250 ps
76 ps
50
/ 500
SW contr.
SYNC
Sync signal on front
panel.
4 ns
1.2 ns
50
/ 500
1
1. 500
option available on MTCA form factor only.
SW contr.
STARB
Backplane trigger in
PXIe systems.
Requires trigger-timing
card.
4 ns
1.2 ns
PXIe standard
PXIe
standard
MLVDS
Backplane trigger in
MVLDS systems.
4 ns
1.2 ns
MLVDS.4
standard
MLVDS.4
standard
Table 2: External trigger inputs.