19-2233 PC2
2019-02-01
25(50)
ADQ7DC Manual
16-1796 PC2 2019-02-01
25(50)
4.9.1
Setting the level trigger level
The level is given in digital codes:
LEVEL_CODE = ANALOG_TRIGGER_LEVEL / ( ACTUAL_ANALOG_RANGE / 2 ) * 2^15
(8)
The
ACTUAL_ANALOG_RANGE
is the analog full scale range. If a DC-offset is used at the input, add the
DC_OFFSET_CODE
from
.
4.9.2
Level trigger and DBS
The Digital Baseline Stabilizer, DBS, compensates for fluctuations in the baseline. The DC level is then
set by a parameter
DBS_CODE
into DBS. The level trigger is set relative to the analog baseline as:
ANALOG_STEP = ANALOG_TRIGGER_LEVEL - ANALOG_BASELINE
(9)
LEVEL_CODE = DB ANALOG_STEP / ( ACTUAL_ANALOG_RANGE / 2 ) * 2^15
(10)
4.9.3
Controlling noise sensitivity
The level trigger is sensitive to noise since it can detect a step as small as one digital code. This can
cause unwanted triggering. The noise sensitivity is controlled by a hysteresis function,
. After
triggering, the signal has to cross a reset level before it can trigger again.
Setting the reset level far from the trigger level will give a robust trigger.
Setting the reset level close to the trigger level will give a sensitive trigger.
#
DESCRIPTION
USER COMMAND
REF
a
When the signal passes the trigger level, a trigger event is generated and
the first record is captured.
SetupLevelTrigger
b
During the record, incoming triggers are ignored.
c
When the signal passes the trigger level, a trigger event is generated and
the second record is captured. Re-triggering is also controlled by a
hysteresis function.
SetupLevelTrigger
Figure 16: Level trigger introduction.
)
*$
+,
+-,
+,