EZ – Sensor Source
Purpose
Selects the source of environmental sensor data.
Format
EZcdhprst
Default
EZ1111101
Recommended Setting. The default setting for this command is recommended for most applications.
Range
Firmware switches (see description)
Description Setting the EZ-command firmware switches tells the Sentinel V Real-Time to use data from
a manual setting or from an associated sensor. When a switch value is non-zero, the Senti-
nel V Real-Time overrides the manual E-command setting and uses data from the appro-
priate sensor. If no sensor is available, the Sentinel V Real-Time defaults to the manual E-
command setting. The following table shows how to interpret the sensor source switch set-
tings.
Table 29:
Sensor Source Switch Settings
Field
Value = 0
Value = 1
c
Speed Of Sound
Manual EC
Calculate using ED, ES,
and ET
d
Depth
Manual ED
Internal Pressure Sensor
h
Heading
Manual EH
Internal Heading Sensor
p
Pitch (Tilt 1)
Manual EP
Internal Pitch Sensor
r
Roll (Tilt 2)
Manual ER
Internal Roll Sensor
s
Salinity
Manual ES
N/A
t
Temperature
Manual ET
Internal Transducer
Sensor
Example
EZ1111101 means calculate speed of sound from readings, use pressure sensor, transducer
heading, internal tilt sensors, and transducer temperature.