Operator’s Manual Simple TrueArb
Application
37
Waveform Area
This area is divided in two main sections, the
Waveform Graph
area that contains a graphical
representation of the channel waveform and the
Waveform Parameters
area.
The Waveform Graph gives a description of the waveform assigned to the current channel and
sequencer entry. The waveform is described as:
➢
The waveform shape
➢
The waveform duration and frequency
➢
The waveform amplitude
The waveform length in term of number of samples as it was originally defined in the Waveform
List (described below)
The Waveform Parameters area is divided in two parts. The left part contains the vertical parameters of
the selected waveform in terms of Amplitude[Vpp] and Offset[V] or Voltage High[V] / Voltage Low [V].
These two parameters can be specified independently for each sequencer entry and for each output
channel.
In the right part there are the Repetitions[N] and the Entry Length[N] for the selected sequencer entry.
These two parameters are specific for the selected
sequencer entry.
In particular Repetitions[N] is the
number of repetitions of the selected sequencer entry.
These parameters are common to all the channels in the same sequencer entry.
Note: Repetitions[N]=1 means that the waveform is executed only once.
Waveform Graph