Rear Panel Description

The rear panel of the Analyzer has: 

Wide-range AC Connector Module

Power socket

Enclosed 5.0 A 250 V fuse

Sync In and Sync Out

: Allows multiple Teledyne LeCroy analyzers to send 

synchronization and control messages to one another.

Warning!  

For continued protection against fire, replace fuse 
only with the type and rating specified above.

Front Panel Description

The Analyzer has the following features on the front:

[1] Power Switch

 (black): 1 = On and 0 = Off.

[2] LCD Menus

: Allows you to configure Summit T416 and view status.

[3] Navigation Buttons

: Navigate through the LCD menu.

[4] Speed LEDs

: Indicate speed setting the analyzer is using for each 

traffic direction.

[5] Upstream [15:0] LEDs

: Lane activity for upstream lanes. OFF for 

electrical idle, orange for activity with errors detected, green for normal 
operation.

[6] Upstream [15:8] and Upstream [7:0] connectors

: Connection to the 

probe for the capture of upstream direction of the link.

[7] Downstream [15:0] LEDs

: Lane activity for downstream lanes. OFF 

for electrical idle, orange for activity with errors detected, green for 
normal operation.

[8] Downstream [15:8] and Downstream [7:0] connectors

: Connection 

to the probe for the capture of downstream direction of the link.

[9] Trigger IN and Out

: Provide external trigger capabilities. You can 

configure Summit T416 to trigger external equipment using Trigger Out. 
You can use Trigger In to trigger the Analyzer from another device.

 

[10] USB 3.0 Type B Host Machine Connector

: To connect the Analyzer to 

the host machine using a USB connection.

 

[11] Ethernet Port

: 1 GIGE Connectivity allows connection to the analyzer 

through an Ethernet network.

 

[12] Status LED

: Displays the connection status of the link.

     

 Red -  

Error condition. Try using the PCIe Protocol Analysis 
application to clear the error. Please contact the factory 
if this does not work. 

     

Green -  

Configuring system.

     

 Blue -  

Unit ready/connected.

[13] Trigger LED

: Lights when Summit T416 triggers on an event.

[14] DS Ref Clk In:

 Downstream Reference Clock Input SMA.

[15] US Ref Clk In:

 Upstream Reference Clock Input SMA.

LCD Menus

The Summit T416 has a front LCD panel that displays the computer name that 
is currently connected to the system, the unit serial number, and the current 
network configuration. If no computer is connected to the unit, the LCD panel 
displays 

Available.

The LCD panel also allows you to configure network settings for the analyzer. 
Use the 

Up

 and 

Down

 keypad buttons to cycle through the menu. Press the 

Center

 button to select.

Setup Static or Dynamic IP Mode

1.

Use the 

Up

 and 

Down

 buttons to navigate to the 

Set IP Configuration

 

menu.

2.

Press the 

Center

 button to start the setup. 

3.

Use the 

Up

 and 

Down

 buttons to select 

Static

 or 

Dynamic

 

IP configuration. 

4.

If you select the Static IP network configuration, you must specify the 
IP address, Subnet mask, and Gateway. Use the 

Up

 and 

Down

 buttons 

to navigate to the IP address, Subnet mask, or Gateway. 

5.

Press the 

Center

 button to enter edit mode.

6.

In edit mode, use the 

Left 

and 

Right 

buttons to move the blinking 

cursor to any digit.

7.

Use the 

Up

 and 

Down

 buttons to modify the digit. 

8.

After finishing modifications to the IP address, Subnet mask, or 
Gateway, press the 

Center

 button to return to the menu.

9.

After you specify the IP address, Subnet mask, and Gateway, use the 

Up

 and 

Down

 buttons to navigate to the 

Accept and Reboot

 menu 

item. Press the 

Center 

button to apply the new settings and restart the 

analyzer. 

Note:

 To cancel your modifications, select the 

Cancel Changes

 menu 

item.

Warning

!

Do not open the enclosure. No operator 
serviceable parts are inside.

1

5

4

5

3

2

6

7

12

8

13

9 10  11

7

15

14

4

5

Sync Out   Sync In

AC Connector

Unpacking the Analyzer

Inspect the received shipping container for any damage. Unpack the container and account for each of the system components listed on 
the accompanying packing list. Visually inspect each component for absence of damage. 

In the event of damage, notify the shipper and Teledyne LeCroy. Retain all shipping materials for shipper’s inspection.