Front Panel Description

The Analyzer has the following features on the front:

[1] Power Switch

 (black): 1 = On and 0 = Off.

[2] Upstream [7:0] connector

: Connection to the probe for the 

capture of upstream direction of the link.

[3] Upstream [7:0] LEDs

: Lane activity for upstream lanes. OFF 

for electrical idle, orange for activity with errors detected, 
green for normal operation.

[4] Speed LEDs

: Indicate speed setting the analyzer is using for 

each traffic direction.

[5] LCD Menus

: Allows you to configure Summit T3-8 and view 

status.

[6] Navigation Buttons

: Navigate through the LCD menu.

[7] Downstream [7:0] connector

: Connection to the probe for 

the capture of downstream direction of the link.

[8] Downstream [7:0] LEDs

: Lane activity for downstream 

lanes. OFF for electrical idle, orange for activity with errors 
detected, green for normal operation.

[9] Trigger IN and Out

: Provide external trigger capabilities. You 

can configure Summit T3-8 to trigger external equipment 
using Trigger Out. You can use Trigger In to trigger the 
Analyzer from another device.

 

[10] Expansion Slot

: For CATC Sync Expansion Card to allow 

multiple Teledyne LeCroy analyzers to send synchronization 
and control messages to one another.

 

[11] USB Type B Host Machine Connector

: To connect the 

Analyzer to the host machine using a USB connection.

 

[12] Ethernet Port

: 1 GIGE Connectivity allows connection to 

the analyzer through an Ethernet network.

 

LCD Menus

The Summit T3-8 has a front LCD panel that displays the computer 
name that is currently connected to the system, the unit serial 
number, and the current network configuration. If no computer is 
connected to the unit, the LCD panel displays 

Available.

The LCD panel also allows you to configure network settings for 
the analyzer. Use the 

Up

 and 

Down

 keypad buttons to cycle 

through the menu. Press the 

Center

 button to select.

Setup Static or Dynamic IP Mode

1.

Use the 

Up

 and 

Down

 buttons to navigate to the 

Set IP Configuration

 menu.

2.

Press the 

Center

 button to start the setup. 

3.

Use the 

Up

 and 

Down

 buttons to select 

Static

 or 

Dynamic

 

IP configuration. 

4.

If you select the Static IP network configuration, you must 
specify the IP address, Subnet mask, and Gateway. Use the 

Up

 and 

Down

 buttons to navigate to the IP address, Subnet 

mask, or Gateway. 

5.

Press the 

Center

 button to enter edit mode.

6.

In edit mode, use the 

Left 

and 

Right 

buttons to move the 

blinking cursor to any digit.

7.

Use the 

Up

 and 

Down

 buttons to modify the digit. 

8.

After finishing modifications to the IP address, Subnet mask, 
or Gateway, press the 

Center

 button to return to the menu.

9.

After you specify the IP address, Subnet mask, and Gateway, 
use the 

Up

 and 

Down

 buttons to navigate to the 

Accept and 

Reboot

 menu item. Press the 

Center 

button to apply the new 

settings and restart the analyzer. 

Note:

 To cancel your modifications, select the 

Cancel Changes

 menu item.

Warning

!

Do not open the enclosure. No operator 
serviceable parts are inside.

1

6

3

4

5

9

11

10

2

8

7

12

4

5

Rear Panel Description

The rear panel of the Analyzer has: 

Wide-range AC Connector Module

Power socket

Enclosed 5.0 A 250 V fuse

Expansion Slot

: Future use

Warning!  

For continued protection against fire, replace 
fuse only with the type and rating specified 
above.

6

Interposers and Probes

Gen3 Slot Interposer

The Gen3 Slot Interposer is designed for use with the Summit T3-8 
Analyzer and supports lane widths from x1 to x8 at data rates of 2.5 GT/s, 
5.0 GT/s or 8.0 GT/s. A separate interposer is used for each lane width 
that you would like to probe. Lane reducers should NOT be used with the 
Gen3 Slot Interposer. For lane widths up to x8, one Analyzer Y-cable is 
required. 

Gen3 MidBus Probe

The Gen3 MidBus Probe is designed for use with the Summit T3-8 
Analyzer and supports lane widths from x1 to x8 at data rates of 2.5 GT/s, 
5.0 GT/s or 8.0 GT/s.

Warning

!

Interposer has electrostatic-sensitive components. 
Handle only at static-safe work stations.

7

Installing the Software

The PCIe Protocol Analysis operates all of Teledyne LeCroy’s PCI Express Protocol Analyzer and Exerciser products and and should be installed on a 
Microsoft

®

 Windows

®

-based host machine. You need to install the PCIe Protocol Analysis on the host machine before attaching the Analyzer to the system. 

The minimum requirements for the host machine are:

Windows 8 (x86 and x64), Windows Server 2012 (x64), Windows 7 (x86 and x64), Windows Server 2008R2 (x64), Windows XP (x86)

The latest Service Pack available for the Windows OS in use is required.

It is recommended that you use one of the supported 64-bit Windows versions listed above as they allow using more RAM than the 32-bit ones.

This software application may use up to 4GB of the RAM in the host machine. For improved performance of the software, it is recommended that 
16GB of RAM is installed on the host machine. Memory as little as 2GB would still allow the software to function, but would limit its performance and 
user experience.

Storage space of 1GB is required for installing the PCIe Protocol Analysis on the host machine.

8

Содержание Summit T3-8

Страница 1: ...ress Protocol Analyzer Quick Start Guide Components The analyzer package includes the following components Summit T3 8 analyzer system AC Power Cable USB and Ethernet cables PCIe Protocol Suite Software program DVD ROM Quick Start Guide this document Depending on configuration the package may also include One probe cable for x8 recording Slot interposer MidBus Probe kit Please see the Summit T3 8 ...

Страница 2: ...n buttons to modify the digit 8 After finishing modifications to the IP address Subnet mask or Gateway press the Center button to return to the menu 9 After you specify the IP address Subnet mask and Gateway use the Up and Down buttons to navigate to the Accept and Reboot menu item Press the Center button to apply the new settings and restart the analyzer Note To cancel your modifications select t...

Страница 3: ...or Summit T3 8 to operate over a network Note 2 At power on the Analyzer initializes and performs a self diagnostic The results are reflected by messages on the Summit T3 8 LCD display If the LCD display indicates failure call Teledyne LeCroy Customer Support for assistance Note 3 Summit T3 8 is configured at the factory to use dynamic IP setting and will get all required network parameters from t...

Страница 4: ...ne LeCroy Inc All rights reserved Part Number 923222 00 Rev D This document may be printed and reproduced without additional permission but all copies should contain this copyright notice Making Your First Recording Connect Summit T3 8 to the system under test using an interposer or probe and then configure the Recording Options You can then test the Analyzer by creating a short snapshot recording...

Отзывы: