General Safety Information 

Observe generally accepted safety procedures in addition to those 

listed here to avoid personal injury or damage to equipment. The 

overall safety of any system incorporating this accessory is the 

responsibility of the assembler of the system. 

Connect only to grounded instruments. Use only with compatible Teledyne 

LeCroy oscilloscopes that have their BNC input connected to an earth 

ground. Do not connect the probe reference lead to any point which is at a 

potential other than earth ground. 

Connect and disconnect properly. Connect probe to the oscilloscope before 

connecting it to the test circuit. Disconnect the probe input and reference 

lead from the test circuit before disconnecting from the oscilloscope. Do 

not connect/disconnect probes while connected to a voltage source. 

Do not overload. Do not apply any potential to the probe leads that exceeds 

the maximum rating of the probe. Observe all terminal ratings on the 

oscilloscope before connecting. Consult the oscilloscope product manual 

for further ratings information. 

Always comply with the Voltage vs. Frequency Derating Curve. 

Do not remove probe casing. Touching exposed connections may result in 

electric shock. 

Use indoors only within operational environment listed. Do not use in wet or 

explosive atmospheres. Keep product surfaces clean and dry. 

Handle with care. Probe tips are extremely sharp and may puncture skin or 

cause other bodily injury if not handled properly. 

Keep fingers behind the finger guard of probe body and accessories. 

Содержание PP021 Series

Страница 1: ...PP021 1 PP021 2 User Manual PP021 Passive Probe...

Страница 2: ...ne LeCroy documentation materials is strictly prohibited Information contained herein is classified as EAR99 under the U S Export Administration Regulations Export reexport or diversion contrary to U...

Страница 3: ...oduction Teledyne LeCroy PP021 is a passive high impedance voltage oscilloscope probe designed and calibrated for use with general purpose oscilloscopes having an input impedance of 1 M It can be comp...

Страница 4: ...or category ratings of the probe See voltage derating curve on p 3 PollutionDegree PollutionDegree2 OperatingAltitude Upto3000meters Max OperatingTemp 0to50 C Humidity 5 to 95 RH 10 to 30 C 5 to 75 RH...

Страница 5: ...User Manual EAR99 technology subject to restrictions on copyrights page 3 PP021 Input Impedance Profile Voltage vs Frequency Derating Curve...

Страница 6: ...he European Communities IEC EN 61010 031 2015 Safety requirements for electrical equipment for measurement control and laboratory use Part 031 Safety requirements for handheld probe assemblies for ele...

Страница 7: ...t Disconnect the probe input and reference lead from the test circuit before disconnecting from the oscilloscope Do not connect disconnect probes while connected to a voltage source Do not overload Do...

Страница 8: ...on Do not allow any portion of the probe to be submerged at any time Dry the probe and accessory thoroughly before attempting to make any voltage measurement Do not subject the probe to solvents or so...

Страница 9: ...probe is connected to or transferred between oscilloscopes Low Frequency 1 Apply a 1 kHz square wave to the probe or connect to the oscilloscope s calibrator output 2 Adjust the single LF trimmer loca...

Страница 10: ...ghten the tips to avoid damaging the threads Ground Accessories The ground lead and the short collar are attached to the probe at the time of shipping To replace the ground lead with the swivel lead a...

Страница 11: ...Trimmer Tool 1 6 PK5 2 5MM 105 Long Collar 1 7 PK5 2 5MM 106 IC Tip Insulators Black Brown Blue Green Gray 5 8 PK5 2 5MM 107 Tip Insulator 2 5mm 1 9 PK5 2 5MM 108 Probe Tip Cover 1 10 PK5 2 5MM 109 BN...

Страница 12: ...nty Teledyne LeCroy at its option will either repair or replace any assembly returned within the warranty period to the Customer Service Department of an authorized service center However this will be...

Отзывы: