Probe Compensation 

Proper compensation of the probe is required to assure good amplitude 
accuracy in the dynamic portions of the waveform being measured.  LF 
compensation matches the probe to differences in oscilloscope input 
capacitance.  The LF compensation should always be checked and adjusted 
as needed when first connecting a passive probe to the oscilloscope input.  
HF compensation matches time constants within the probe to compensate 
for normal component tolerances.  It is typically not necessary to adjust HF 
compensation unless the probe is being used with an oscilloscope with large 
differences in input characteristics than the oscilloscope model it was 
designed for. 
LF compensation is performed by connecting the input of the probe to a low 
frequency square wave, such as the oscilloscope calibrator signal set to  
1 kHz. The compensation is adjusted by rotating the adjustment accessible 
through the small hole in the center of the housing near the BNC connector.  
Use the tool supplied with the probe for this adjustment. 

 

Undershoot 

 

Overshoot 

 

Correct adjustment 

Should HF compensation be required, access the 
adjustments by sliding the black plastic cover off 
the compensation housing near the BNC con-
nector. A pulse generator with low overshoot and a 
300 ps risetime is the required signal source, along 
with a set of attenuators. The probe must be 

connected to a terminated probe tip to BNC 
adapter. 
Some overshoot and ring will be present at some 
settings of V/Div.  Adjust both trimmers for the 
overall best response on all ranges.  

 

Typical optimum HF 

adjustment 

Содержание PP008

Страница 1: ...PP008 1 PP008 2 Operator s Manual PP008 Passive Probe...

Страница 2: ...ne LeCroy at its option will either repair or replace any assembly returned within its warranty period to the Customer Service Department or an authorized service center However this will be done only...

Страница 3: ...erstood and met DOUBLE INSULATION PROTECTIVE EARTH TERMINAL Precautions To avoid personal injury or damage to property review and comply with the following safety precautions Use product only as speci...

Страница 4: ...s case or touching exposed connections may result in electric shock Use only within operational environment listed Use indoors only Do not use in wet or explosive atmospheres Keep product surfaces cle...

Страница 5: ...ly Do not operate where conductive pollutants may be present Maximum input voltage derates when frequency response is approximately 30 kHz or higher See voltage derating curve on p 4 Specifications El...

Страница 6: ...to 71o C Altitude Operating Up to 2000 m 6560 ft Humidity Operating 5 to 85 RH to 30 C 5 to 65 RH 30 C to 40 C 5 to 45 RH above 40 C General Characteristics Cable Length 1 3 m 4 3 ft Weight probe only...

Страница 7: ...er followed by the description PK701 Basic Adapter Kit replaces the common standard accessories with Qty 2 each of 1 2 4 5 and 14 plus Qty 1 adjustment tool 3 PK702 Advanced Adapter Kit contains a lar...

Страница 8: ...PK007 005 Spring Tip 0 5 mm PK007 007 Insulating Cap PK007 008 IC Cap 0 5 mm pitch green PK007 009 IC Cap 0 65 mm pitch blue PK007 010 IC Cap 0 8 mm pitch gray PK007 011 IC Cap 1 0 mm pitch brown PK0...

Страница 9: ...18 19 20 PK007 017 Single Lead Adapter PK007 018 Dual Lead Adapter PK007 022 Adapter 2 mm plug PK007 023 Adapter 4 mm plug 21 22 23 PK007 019 Pico Hook PK007 020 Micro Clip Long PK007 021 Micro Clip S...

Страница 10: ...ator Clip 28 29 30 PK007 026 Ground Lead with mini clip PK007 027 Ground Lead 8 mm socket PK007 028 Ground Lead 2 mm plug 31 32 PK007 029 Ground Lead 4 mm plug PK007 030 High Frequency Compensated Gro...

Страница 11: ...ther maintenance and component replacement should be referred to qualified personnel Cleaning The outside of the probe should cleaned with a soft cloth dampened with either deionized distilled water o...

Страница 12: ...for LF compensation is performed by connecting the input of the probe to a low frequency square wave such as the oscilloscope calibrator signal set to 1 kHz The compensation is adjusted by rotating th...

Страница 13: ...Select the proper tip for your application needs To replace tips carefully grip the outermost portion and pull straight out along the axis of the probe using needle nose pliers Do not grip the plastic...

Страница 14: ...cted to utilization points socket outlets and similar of the low voltage mains installation Pollution Degree 2 operating environment where normally only dry non conductive pollution occurs Conductivit...

Страница 15: ......

Страница 16: ...928596 00 Rev A July 2017...

Страница 17: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Teledyne LeCroy PP008 1...

Отзывы: