ZS4000 High-Impedance Active Probe
20
B. LF Attenuation Accuracy
LF Attenuation Accuracy Setup
1.
Disconnect the BNC tee at the power supply from the dual banana plug
adapter. Connect the BNC tee to the output of the function gen erator. (Use
a 50
Ω
termination if the function generator requires such a load.)
2.
Carefully insert the Straight Tips (supplied in accessory kit) into the sockets
of the probe head. Attach the red lead of the mini-grabber to the signal
input and the black lead to the ground input of the probe head.
3.
Connect the BNC tee to the output of the function generator. (Use a 50
Ω
termination if the function generator requires such a load).
4.
Attach a BNC cable to the unused female port of the BNC tee and connect
a dual banana plug adapter to the other end of the cable and plug the dual
banana plug adapter into the DMM input. Be sure the side of the banana
plug adapter corresponding to the BNC shield (marked "GROUND") is
connected to the
LOW
or
COMMON
input of the DMM.
5.
Set the DMM to read AC volt and set the range to measure 5.0 Vrms.
Содержание Everywhereyoulook ZS4000