The plot below shows a comparison of both response modes (at their 
maximum bandwidth setting). 

0

5

10

15

20

25

30

35

40

Frequency (GHz)

-6

-5

-4

-3

-2

-1

0

1

Response (dB)

OE6250G-M: Nominal Compensated Response

4th-order Bessel-Thomson response
Flatness-opimized response

 

Optical Wavelength 

The optical wavelength of the OE6250G-M can be selected between  
830 nm and 1600 nm in increments of 10 nm.  
The OE6250G-M is calibrated at 850 nm, 1310 nm, and 1550 nm.  
Other wavelength responses are interpolated/extrapolated from the calibrated 
values. 

Содержание Everywhereyoulook OE6250G-M

Страница 1: ...User Manual OE6250G M Optical Electrical Converter...

Страница 2: ...unauthorized duplication of Teledyne LeCroy documentation materials is strictly prohibited Teledyne LeCroy is a registered trademark of Teledyne LeCroy Inc Other product or brand names are trademarks...

Страница 3: ...ures 1 Compatibility 1 Safety 2 Operation 3 Connecting the Unit 3 Controlling from the Oscilloscope 4 Care and Maintenance 6 Cleaning the Converter Body 6 Cleaning and Connecting Optical Fibers 6 Cali...

Страница 4: ...authorized service centers within this period However in order to do this we must first examine the product and find that it is defective due to workmanship or materials and not due to misuse neglect...

Страница 5: ...ce the OE6250G M becomes an integrated part of the oscilloscope The Optical wavelength dark calibration and reference receiver functions are all controlled from the oscilloscope s touch screen The osc...

Страница 6: ...always clean the optical connector and ensure the facet is free from dirt or impurities See the instructions on p 6 Do not install or terminate fibers while a light source is active Care must be taken...

Страница 7: ...e and pushing until you hear a click 2 Connect one end of the 2 92 mm cable to the 2 92 mm connector on the OE6250G M and the other to the 2 92 mm connector on the oscilloscope 3 Clean and connect the...

Страница 8: ...set signal without any light present in the optical receiver The Set to Zero function resets the Dark Calibration value to zero The Dark Level field allows manual entry of the Dark Level offset this w...

Страница 9: ...dB OE6250G M Nominal Compensated Response 4th order Bessel Thomson response Flatness opimized response Optical Wavelength The optical wavelength of the OE6250G M can be selected between 830 nm and 160...

Страница 10: ...it to the port If the fiber is dirty clean it as follows 1 Gently wipe the fiber end with a lint free swab dipped in isopropyl alcohol 2 Use compressed air to dry it completely 3 Visually inspect the...

Страница 11: ...he power connector is plugged in and locked and that the 2 92mm cable is connected Service Defective products must be returned to a Teledyne LeCroy service facility for diagnosis and repair or replace...

Страница 12: ...requirements to Directives 2012 19 EU on Waste Electrical and Electronic Equipment WEEE For more information about proper disposal and recycling of your Teledyne LeCroy product visit teledynelecroy c...

Отзывы: