88
Teledyne API T500U CAPS NO
2
Analyzer with NumaView™ Software
083730510A DCN8172
TIME ZONE CHANGES
There is an option to change between 12-hour and 24-hour format in the Setup>Vars menu
(System Time Format). Effectively changing the Time Zone requires a specific procedure
as follows:
1. In Setup>Instrument>Date/Time Settings select the applicable Time Zone.
2. Allow adequate time for the selected Time Zone to be properly accepted.
3. Verify: return to Home page then back to the Date/Time Settings page, and
check that the selected Time Zone is now highlighted.
4. Without making any other changes, power OFF the instrument and power ON
again.
5. Once restarted, return to the Date/Time Settings page where the newly
selected Time Zone should be highlighted. (If not, it means that not enough
time had passed for the instrument to accept the change before the power was
cycled OFF).
6. After the Time Zone is implemented first (Steps 1 through 5), then other
changes to the date and/or time can be made, and recycling the power is not
necessary.
Figure 5-4. Time Zone Change Requirements
Содержание T500U
Страница 2: ......