Teledyne API – Model T300/T300M CO Analyzer
Calibration Procedures
213
Offset for
CO REF
signal
SAMPLE
RANGE=50.0 PPM CO= XX.XX
<TST
TST> CAL
SETUP
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
DIAG
SIGNAL I/O
PREV
NEXT
ENTR
EXIT
SETUP X.X
ENTER PASSWORD
8
1
8
ENTR
EXIT
Continue pressing
NEXT
until ...
DIAG OPTIC
DARK CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG DARK
CO DARK CALIBRATION
VIEW
CAL
EXIT
DIAG DARK
REF DARK OFFSET: 0.0mV
EXIT
Offset for
CO MEAS
signal
DIAG DARK
MEAS DARK OFFSET: 0.0mV
EXIT
DIAG DARK
DARK CAL 1% COMPLETE
EXIT
DIAG DARK
DARK CALIBRATION ABORTED
EXIT
Calibration runs automatically
9.6.2.
PRESSURE CALIBRATION
A sensor at the sample chamber outlet continuously measures the pressure of the sample
gas. These data are used to compensate the final CO concentration calculation for
changes in atmospheric pressure and is stored in the CPU’s memory as the test function
PRES
(also viewable via the front panel). See also Section 5.9.6.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
This calibration must be performed when the pressure of the sample gas
is equal to ambient atmospheric pressure. Before performing the
following pressure calibration procedure, disconnect the sample gas
pump and sample gas-line vent from the rear panel sample gas inlet.
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...