5. 

INSTRUMENT ERROR CODES 

Table 5-1 shows the instrument error codes, their causes, and corrective action. 

Table 5-1. Instrument Error Codes 

Error 

Code

 

Cause

 

Corrective Action

 

Too much light - out of design range.   The 

instrument may have been zeroed while 

SiF4 was present. 

Purge and zero the instrument. 

No light detected. 

Purge and zero the instrument.  If error persists after 

zeroing: 

  Do not open the unit without prior permission 

from TAPI.   

  Contact TAPI immediately for service 

instructions.   It may be possible to repair your 

unit in situ. 

  If you do remove the unit, check the KF16 

ports for powder contamination.  Powder 

contamination indicates presence of particulate 

matter or a deposited film.  Make sure that 

provisions are in place to prevent particulate 

matter or film deposition onto the surfaces of 

the absorption cell.  

Too much light - signal cannot be 

measured.   The instrument may have been 

zeroed while SiF4 was present. 

Purge and zero the instrument. 

Instrument warming up. 

Will clear automatically three minutes after the 

instrument boots up. 

A zeroing request was invalid because the 

instrument was warming up.  The instrument 

did not honor the zeroing request. 

Wait for the three minute warm-up period to complete, 

then re-zero the instrument. 

Insufficient light. 

Same as for Error Code 3. 

Timing error. 

Purge and zero the instrument.  Contact TAPI if the 

problem occurs repeatedly. 

Missed reading. 

Purge and zero the instrument.  Contact TAPI if the 

problem occurs repeatedly. 

 

Содержание OPTI-Sense 5200

Страница 1: ...r TELEDYNE API TAPI 9970 CARROLL CANYON ROAD SAN DIEGO CALIFORNIA 92131 1106 USA Toll free Phone 1 800 324 5190 Phone 1 858 657 9800 Fax 1 858 657 9816 Email api sales teledyne com Website http www te...

Страница 2: ......

Страница 3: ...OF COPYRIGHT 2022 Teledyne API TAPI All rights reserved TRADEMARKS All trademarks registered trademarks brand names or product names appearing in this document are the property of their respective own...

Страница 4: ...ion or proper tools could result in damage to the part s and or the instrument Technician Symbol All operations marked with this symbol are to be performed by qualified maintenance personnel only Elec...

Страница 5: ...gnements sp cifiques MISE EN GARDE Surface chaude Ne pas toucher Toucher certaines parties de l instrument sans protection ou sans les outils appropri s pourrait entra ner des dommages aux pi ces ou l...

Страница 6: ...vided but not manufactured by TAPI is warranted and will be repaired to the extent and according to the current terms and conditions of the respective equipment manufacturer s warranty PRODUCT RETURN...

Страница 7: ...the careful and effective use of the instrument and related pertinent information ATTENTION COULD DAMAGE INSTRUMENT AND VOID WARRANTY This special notice provides information to avoid damage to your i...

Страница 8: ...L INSTALLATION 9 3 ELECTRICAL CONNECTIONS 11 RS 232 Connector 11 I O Connector 11 Fault DO Specifications 12 Over Voltage Protection 12 4 SYSTEM OPERATION 13 Zeroing the OPTI Sense 13 5 INSTRUMENT ERR...

Страница 9: ...iod is completed CAUTIONS Please review these instructions carefully to ensure proper installation operation and care of the system to minimize the risk of personal injury or product damage due to imp...

Страница 10: ...second Analog scaling Filter type 1 10 VDC corresponds to 0 0 0 2 units of absorbance 10Watts factory set Leak Testing Vacuum Leak Tight Cell Gas Connection KF16 300 Series stainless steel Pressure Ra...

Страница 11: ...ENTION Ensure adequate space to access connectors and to remove cover when necessary for servicing Ensure adequate clearance to allow free air flow for cooling and to avoid high heat generating equipm...

Страница 12: ...is also acceptable Figure 2 2 Inlet right and Outlet left Connectors The flow of gas through the unit should be interrupted when there is a possibility of particulate matter soiling the windows or lo...

Страница 13: ...nnected to the COM1 port of the computer running the GUI software I O CONNECTOR Table 3 1 below shows the I O connector pinout Table 3 1 I O Connector Pinout Pin Connection Notes 1 N C 2 Analog Signal...

Страница 14: ...automatically when they no longer affect instrument data For example the three minute instrument warm up period is indicated as an error to allow detection of accidental power disruption the Fault DO...

Страница 15: ...ow the line must be held low for at least 200 ms Operating Voltage HIGH 24VDC 10 Operating Voltage LOW 0VDC Nominal 4VDC Max Zeroing will take approximately five seconds during normal operation Zeroin...

Страница 16: ...r contamination indicates presence of particulate matter or a deposited film Make sure that provisions are in place to prevent particulate matter or film deposition onto the surfaces of the absorption...

Страница 17: ...chnical Support department for any technical assistance needed Teledyne API Technical Support 9970 Carroll Canyon Road San Diego California 92131 1106 USA Toll free Phone 1 800 324 5190 Phone 1 858 65...

Отзывы: