42
Teledyne API Model 480H O
3
Monitor User Manual
09421A DCN8343
7. MAINTENANCE
WARNING – ELECTRICAL SHOCK HAZARD
The operations outlined in ths section are to be performed
by qualified maintenance personnel only!
CAUTION – RISK OF PERSONAL INJURY: HAZARDOUS GAS
Prior to opening the front panel for service of any kind, ensure first
that the area is well-ventilated before slowly opening the panel and
allowing gradual dissipation of any HAZARDOUS GAS that may
have accumulated within the enclosure.
CAUTION – RISK OF PERSONAL INJURY: UV RADIATION
Do not look at the UV LED while the unit is operating. UV light can
cause eye damage. Always use safety glasses made from UV
blocking material. (Generic plastic glasses are not adequate).
7.1. MAINTENANCE SCHEDULE
Table 7-1 below outlines the suggested maintenance procedures and intervals for
ensuring the 480H continues to operate accurately and reliably. These intervals are
based on continuous (24 hours a day – 7 days a week) operation. These intervals
may be lengthened for intermittent operation.
Table 7-1. Maintenance Schedule
Maintenance Item
Recommended Interval
Section
Replace particulate filter
element
6 months
Replace UV LED
When UV LED intensity drops to a
level that triggers the Check Lamp
status, and the Sensor OK status
indicates an ERROR.
Replace Auto-Zero Valve
annually
Rebuild cells, replace
o-rings
3 years
Return to
factory
Cleaning
As needed.
Содержание Everywhereyoulook 480H
Страница 2: ...E IS INTENTIONALLY LEFT BLANK...
Страница 6: ...iv Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 This page intentionally left blank...
Страница 12: ...10 Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 This page intentionally left blank...
Страница 17: ...09421A DCN8343 Teledyne API Model 480H O3 Monitor User Manual 15...
Страница 40: ...38 Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 Figure 6 1 RS 232 Signal Configurations...