Portable Trace Oxygen Analyzer
Operation
Teledyne Analytical Instruments
21
3.2.9 LOG INTV Screen (Optional)
This screen is for use with the data logger option only. It indicates the
time interval between data samples taken by the data-logger. The interval
can be set between 1 second (00m 01sec) to 60 minutes (60m 00sec) in
increments of 1 second.
To change the interval between data samples:
1.
Use the UP/DOWN keys to navigate to the
LOG INTV
screen. Then press ENTER to enter the setup screen.
2.
Use the UP /DOWN keys to change the interval value. The
value of the seconds will increase or decrease followed by
the minutes. Press ENTER to accept the displayed value and
return to the
LOG INTV
screen.
3.2.10 LOG RESET & START Screen (Optional)
Once a log interval has been input, the
LOG RESET & START
screen is used to start the data-logger.
The data logger has a 3200 record capacity and each record uses 10
bits of data. The data set is retained until it is reset by using the START
RESET menu again. The data set can be downloaded using the LOG
TRANSMIT screen (see Section 3.2.11).
To start the data logger:
1.
Use the UP/DOWN keys to navigate to the
LOG RESET &
START
screen and press ENTER.
2.
When the screen begins to blink press ENTER again to start
the data logger. Once the data logger has started, the screen