![Tektronix WFM90D Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/tektronix/wfm90d/wfm90d_quick-start-manual_1078268020.webp)
Specifications
1- 2
WFM90D and WFM91D Service Manual
Characteristics Tables
The tables use the following abbreviations:
Performance Requirement (Req).
All performance requirements in the specifica-
tion tables are identified with
Req
, and can be assured by completing the
Performance Check Procedure located in the service manual. Allow a warm-up
time of 20 minutes.
Performance requirements in the electrical specifications are valid over an
ambient temperature range of +20
°
C to +30
°
C. The Performance Requirement
tolerances listed in the Electrical Specification are doubled over the temperature
range of 0
°
C to +40
°
C, unless otherwise specified. Test equipment used to
verify Performance Requirements must be calibrated and working within the
limits specified in the service manual Equipment Required List.
Reference Information
.
Information that amplifies a performance requirement or
is of special importance is indicated by
REF
. There is no need to check these
items to a specific tolerance.
Electrical Specifications
Table 1- 1: Vertical deflection system
Category
Description
Frequency Response
Analog Input
REF:
Specifications apply for full screen height video input signal, with VARIABLE GAIN off.
REQ:
Flat Filter 1 V Full Scale:
50 kHz to 6 MHz within 2% of response at 50 kHz.
REQ:
Flat Filter X5 Gain:
50 kHz to 6 MHz within 5% of response at 50 kHz.
REF:
Low Pass Filter:
≥
40 dB attenuation at 3.58 MHz NTSC, 4.43 MHz PAL.
REF:
Response at 15 kHz does not vary between FLAT and LUM (low pass) filters by more than 1%.
Digital Input
REF:
Specifications apply for full screen height video input signal, with VARIABLE GAIN off.
REQ:
Flat Filter 1 V Full Scale:
50 kHz to 4.5 MHz
±
4%.
REQ:
Flat Filter X5 Gain:
50 kHz to 4.5 MHz
±
5%.
REF:
Low Pass Filter:
≥
40 dB attenuation at 3.58 MHz NTSC, 4.43 MHz PAL.
REF:
Response at 15 kHz does not vary between FLAT and LUM (low pass) filters by more than 1%.
Содержание WFM90D
Страница 4: ......
Страница 14: ...Service Safety Summary x WFM90D and WFM91D Service Manual...
Страница 17: ...Specifications...
Страница 18: ......
Страница 29: ...Operating Information...
Страница 30: ......
Страница 44: ...Functional Overview 2 14 WFM90D and WFM91D Service Manual...
Страница 57: ...Theory of Operation...
Страница 58: ......
Страница 62: ...Theory of Operation 3 4 WFM90D and WFM91D Service Manual...
Страница 63: ...Performance Verification...
Страница 64: ......
Страница 82: ...Performance Verification Digital Section 4 18 WFM90D and WFM91D Service Manual...
Страница 83: ...Adjustment Procedures...
Страница 84: ......
Страница 88: ...Adjustments 5 4 WFM90D and WFM91D Service Manual...
Страница 98: ...Adjustment Procedures Digital Section 5 14 WFM90D and WFM91D Service Manual...
Страница 99: ...Maintenance...
Страница 100: ......
Страница 130: ...Remove and Replace 6 30 WFM90D and WFM91D Service Manual...
Страница 152: ...Troubleshooting Procedures 6 52 WFM90D and WFM91D Service Manual...
Страница 153: ...Options...
Страница 154: ......
Страница 157: ...Replaceable Mechanical Parts...
Страница 158: ......
Страница 166: ...Replaceable Parts List 8 8 WFM90D and WFM91D Service Manual...
Страница 167: ...Diagrams...
Страница 168: ......
Страница 170: ...Diagrams 9 2 WFM90D and WFM91D Service Manual...