background image

Release Notes

Calibration and

Performance Veri

fi

cation

Before entering Calibration, connect any supported SDI signal to the SDI A

input, select Input A, verify lock, then disconnect the SDI signal and start
Calibration.

On some units, the calibration screen might not be displayed after selecting

CONFIG > Utilities > Calibration. If this occurs, press the STATUS button to
display the calibration screen.

If a clean waveform is not seen when Eye Gain Adjust is selected, then

remove the calibration signal and connect an SD SDI signal to the selected
SDI input. After the SDI eye diagram is stable, remove the SDI signal and
reconnect the calibration signal.

After calibration is complete, instrument power must be cycled to restore

normal operation.

Display Distortion

The audio phase display is optimized for a 4 x 3 monitor. When an external 16 x 9
XGA monitor is used, the X-Y axes of the audio phase display are not 90 degrees
apart and the waveform is stretched somewhat horizontally.

Capture and Freeze

Capture Buffer is designed to capture digital video only, and not analog video.

You can select Capture Buffer while monitoring a composite video input,

but there will be no effect because the Capture Buffer mode does not apply
to composite analog signals. Use the Capture Freeze mode when monitoring
composite analog signals.

Short duration trigger events are captured, but “frozen” status displays such as

video session may not update in time to re

fl

ect the event.

Captured format must match input. When viewing capture buffer data, ensure

a signal with the same format is on the selected input.

Capture mode works when at least one tile is one of the following: waveform,

vector, gamut, or picture.

Use Freeze instead of Buffer to capture physical layer characteristics.

Selecting to delete the capture buffer from the Remote UI does not delete

the capture buffer.

Manipulating Capture data (such as changing display types or changing to

full screen) can introduce spurious errors in the captured status and session
screens.

Dolby Bar Labels

Depending on the current state of the input, surround channels may be indicated
by numbers 1-10, rather than by labels L, R, C, and so on. The channels, if
present in the AUDIO display, allow the Channel Labels to be present in both
the audio input/output maps in the CONFIG menu and in the audio session. If
channels are not present (input shifts from 5.1 to Stereo or downmixes are turned

2

Waveform Monitors Release Notes

Содержание WFM6120

Страница 1: ...WFM6120 WFM7020 WFM7120 WFM6100 Opt MB WFM7000 Opt MB and WFM7100 Opt MB Waveform Monitors ZZZ Release Notes xx www tektronix com P077022600 077 0226 00...

Страница 2: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Страница 3: ...you are viewing data types in the Anc Data Inspector display with Watchlist enabled while in simultaneous input mode SIM Unknown ancillary data types are sometimes reported when inputs are switched Y...

Страница 4: ...a composite video input but there will be no effect because the Capture Buffer mode does not apply to composite analog signals Use the Capture Freeze mode when monitoring composite analog signals Sho...

Страница 5: ...tile shows the Ls and Rs levels attenuated by 3 dB instead of displaying at full amplitude for this channel mode listening mode combination When in Channel Mode 3 0 and 3 Stereo Listening Mode the Su...

Страница 6: ...to lock to an NTSC reference removing the external reference signal leaves the waveform stable but shifted on the screen Note that the SDI Input Readout can be used to determine if the waveform is loc...

Страница 7: ...de but not simultaneous input mode SDI Status Display When the input signal is missing the Cable Length meter the Cable Length readout and the Cable Loss readout on the SDI Status display indicate zer...

Страница 8: ...G is on HD formats are unaffected USB Always use the Unmount USB menu function to safely remove the USB memory device from the USB port You risk permanent loss of any files saved on the USB device if...