VX1411 MainFrame Instruction Manual
5–1
Performance Verification
Use this chapter to verify that the VX1411 Mainframe operates properly. The
checks consist of installing a 73A-850 extender card into slot 0 of the main-
frame. Check the power supply voltages with a DMM.
Equipment Required
You will need the following equipment to complete the performance verification
checks in this chapter:
H
Digital Voltmeter (Tektronix DM250 series digital voltmeter)
H
Extender card, 73A-850.
DC Voltage Regulation
The DC voltage regulation can be verified by following the steps listed under
Functional Check beginning on page 1–22.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com