Getting Acquainted with Your Instrument
2.
Attach the Sync Pickoff BNC connector
to a BNC-to-minigrabbers connector
(Pomona Electronics model number
3789).
3.
Connect the minigrabbers to the PROBE
COMP connector.
4.
Select CH1 and CH4. Push
FastAcq
.
Push
AUTOSET
.
Both the CH1 and CH4 waveforms
should be displayed. It is normal for the
waveform to show tilt, caused by the
75
Ω
Sync Pickoff termination loading
the PROBE COMP output signal.
5.
Compare the Channel 1 and 4
waveforms. If they do not have the
same shape, adjust the Sync Pickoff
compensation (the adjustment is
accessed through the side of the small
housing).
Properly compensated
Under compensated
22