PREVENTIVE MAINTENANCE
Calibration
The Type CA Plug-In Unit is
designed for
maximum
stability
and should not require fre
quent recalibration. However, to insure the
accuracy of
measurements, we
suggest that you
recalibrate
the instrument
after
every 500-
hour
period
of operation (or
every six months
if
the
unit
is used
intermittently).
A complete
step-by-step
procedure for
recalibrating the
unit
and checking its operation is given in the
Calibration section
of this manual. The
accuracy
of
measurements
made with the Type CA Unit
depends
not only on the accuracy of the Type
CA
Unit
calibration but on the associated
oscilloscope
calibration as well.
Therefore, it
is essential that the oscilloscope be maintained
in
proper calibration.
Visual
Inspection
Many
potential
and existent troubles
can be
detected by a
visual inspection of the unit,
for
this
reason, you should
perform a com
plete visual check
every time the instrument
is
recalibrated or repaired. Apparent defects
may include loose or broken
connections, dam
aged connectors, improperly seated
tubes,
scorched
or
burned parts, or broken terminal
strips,
as
well
as
many other troubles. The
remedy
for these troubles is
readily apparent
except
in the case
of heat-damaged parts.
Damage
of parts
due to heat is often the direct
result
of
other,
less
apparent
troubles in the
circuit.
It
is essential that you determine
the
cause of overheating before replacing the
damaged parts to prevent damage to the new
components.
COMPONENT
REPLACEMENT
The procedures for replacing
most parts
in
the
Type
CA
Unit
are obvious. Detailed
instructions for their removal are therefore
not
required.
Other components, however, can
best
be removed if a
definite procedure is
followed
or
if
certain
precautions are
taken.
Additional
information for the replacement
of
some
of these parts is contained in the following
paragraphs.
Because
of the nature
of the
instrument, replacement
of certain parts will
require
that you recalibrate portions
of
the
instrument to insure proper operation. Refer
to
the
Calibration section of this manual.
Tubes
Care should be taken
both in preventive and
corrective
maintenance that tubes
are
not
replaced unless
they are actually
causing a
definite circuit
malfunction.
Many times during
routine
maintenance it will be necessary
for
you
to
remove tubes from their sockets. It is
important
that these
tubes
be returned to the
same sockets unless they are actually defective.
Needless replacement or switching of tubes
will
many times result in
unnecessary
recalibration of the instrument. If tubes
do
require
replacement, it
is recommended
that
they
be replaced by previously checked
high-
quality tubes.
Switches
Methods
for removal of defective
switches are,
for
the
most part, obvious and
only a normal
amount
of
care is required. Single
wafers
are
normally not replaced on the switches
used
in the Type CA
Unit, and if
one
wafer
is
defective
, the entire switch should
be
replaced.
Switches
may
be
ordered from Tektronix
either
wired or
unwired as
desired.
Soldering
and
Ceramic
Strips
Many
of the components
in your Tektronix
instrument
are mounted on ceramic
terminal
AA
Maintenance
- Type
CA
4-1