SECTION  5

 

CALIBRATION

General  Information 

PRELIMINARY  INSTRUCTIONS

The  Type  TU-7  does  not  require  frequent  recalibration, 

however,  the  calibration  should  be  checked  at  regular  inter­

vals  to  insure  that  it  is  operating  properly  and  accurately. 

In  addition,  calibration  should  be  checked  after  tubes  or 

transistors  have  been  replaced  or  repairs  have  been  made. 

A  complete  procedure  is  provided  in  this  section  for  checking 
and  adjusting  the  Type  TU-7.

In  the  instructions  that  follow,  the  steps  are  arranged 

in  convenient  sequence  to  avoid  unnecessary  repetition. 
Individual  steps  can  be  performed  out  of  sequence,  but  the 
equipment  connections  and  control  settings  in  previous  steps 
will  need  to  be  noted.  If  the  DRIVE  BAL  control  (step  12) 

is  adjusted,  the  PULSE  DC  LEVEL  adjustment  (step  13)  must 
be  checked.

Equipment  Required

The  following  items  of  equipment  or  equivalent  are  re­

quired  for  a  complete  calibration  of  the  Type  TU-7.

1.  Indicator  oscilloscope,  Tektronix  530-,  540-,  or  550- 

Series.  Must  operate  properly  and  be  calibrated  for 

vertical  gain  and  sweep  timing.

2.  Test  oscilloscope,  laboratory-type.  Dc  to  at  least  1  me 

bandpass,  ac  and  dc  input  coupling,  0.02  to  20  volfs/cm 

input  deflection  factor.

3.  1 OX probe,  for  test  oscilloscope.  High  impedance  input.

4.  42-inch  50-ohm  coaxial  cable  with  BNC  connectors 

[Tektronix  part  no.  012-057).

5.  Amplitude  calibrator. 

1  kc  square-wave  output,  am­

plitudes  from  at  least  0.5  volt  to  100  volts  peak-fo-peak, 
accuracy  within  3%.  The  oscilloscope  Calibrator  may  be 

used.

1.  Lay  the  indicator  oscilloscope  (item  1  under  "Equipment 

Required” )  on  its  right  side.  If  rackmounfed,  leave  oscillo­

scope  upright.

2.  Remove  the  oscilloscope  bottom  panel  and  left  side  panel.

3.  Connect  the  test  instruments  to  the  power  line.

4.  Turn  on  all  test  instruments  except  the  indicator  oscillo­

scope.

5.  Set  the  indicator  oscilloscope  front-panel  controls  as  fol­

lows:

Horizontal  Display 

Time  Base  A

Time/Cm 

1  mSEC

Triggering 

+  Internal,  Ac

6.  Set  the  sampling  oscilloscope  controls  as  follows: 

Time/Cm 

1  nSEC

Triggering 

-(-Internal

7.  Set  the  Type  TU-7  controls  as  follows:

NOTE

Do  not  insert  the  Type  TU-7  into  the  ind icato r 
oscilloscope  until  instructed  later  in  the  procedure.

TEST  FUNCTION 

VARIABLE

VERTICAL  POSTION 

AMPLITUDE 

REPETITION  RATE

LOW  LOAD 

Centered 

Centered 

Centered 

MED

PROCEDURE

6.  Dc  voltmeter.  20,000  ohms/volt  impedance,  0.5%  ac­

curacy  at  67.5-voIt  reading.

7.  Ohmmeter  or  multimeter.  Resistance  scale  of  1  ohm,  10 

ohms,  1,000  ohms,  and  100,000  ohms,  3%  accuraccy.

8.  Oscilloscope,  sampling-type,  dual-trace.  50-ohm  inputs 

with  GR  connectors,  at  least  10  to  lOOmv/cm  input  de­

flection  factor,  bandpass  to  1  gc  [gigacycle  —  I0 9cps), 
capable  of  operating  in  differential  mode.  This  instru­

ment  and  the  following  two  items  are  required  only  for 

checking  the  risetime  of  the  Pulse  Generator  output.

9.  Two  50-ohm  coupling  capacitors  (874-K),  with  GR  con­

nectors,  for  sampling  oscilloscope.  (Tektronix  part  no. 
017-028).

10.  16-pin  plug-in  extension,  specially  modified  for  checking 

Pulse  Generator  output.  Construction  of  the  extension  is 

illustrated  in  Fig.  5-1.

1.  Check  Resistances

NOTE

This  step  is  required  only  if  the  instrument  has

 

been  repaired  or  has  been  malfunctioning.

a.  With  the  aid  of  Table  4-1,  measure  resistance  from 

chassis  ground  to  the  interconnecting  plug  pins.

2.  Check  Protection  Diode— Pulse  Generator

a.  Set  the  ohmmeter  to  the  1-ohm  scale  and  check  the 

in-circuit  resistance  across  D54  for  approximately 

5  ohms  in  ore  direction;  infinite  in  the  other  direc­
tion.  See  Fig.  5-2  for  location  of  D54.

Check  + 2 2 5 V   Output

a.  Insert  the  Type  TU-7  into  the  indicator  oscilloscope.

5-1

Содержание TU-7

Страница 1: ...MANUAL Serial N um ber p Tektronix Inc S W Millikan Way P O Box 500 Beaverton Oregon Phone Ml 4 01 61 Cables Tektronix 070 407 36...

Страница 2: ...ctly to the fie ld there fore a ll requests fo r repairs and replace ment parts should be directed to the Tek tro n ix Field O ffice or Representative in your area This procedure w ill assure you the...

Страница 3: ...structions Section 3 C ircuit Description Section 4 M aintenance Section 5 C a lib ra tio n Section 6 Parts List D iagram A list o f abbreviations and sym bols used in this m anual w ill be found on p...

Страница 4: ...TU 7 PLUG IN TEST UNIT SERIAL VERTICAL POSITION AMPLITUDE VARIABLE LOAD POSITIONS ONLY test f u n c t io n HIGH LOAD PULSE REPETITION rate MED LOW LOAD PULSE EXT INPUT PORTLAND OREGON U S A TEKTRONIX...

Страница 5: ...cope vertical amplifier ALTERNATE Checks operation of the alternate mode synchronizing circuits in the oscilloscope Also permits checking for proper sweep slaving in oscilloscopes having two time base...

Страница 6: ...ing external signals to the oscilloscope vertical system through the Type TU 7 Useful for applying the calibrator signal when setting the gain of the oscillo scope vertical amplifier and for inserting...

Страница 7: ...to center the display 4 Adjust the oscilloscope Vertical Gain control for exactly 4 cm vertical distance between the two traces Keep the display centered vertically on the crt while making this adjus...

Страница 8: ...5 Sec Cm 3 Adjust the oscilloscope triggering controls for a stable chopped waveform display Fig 2 2 4 Set the oscilloscope Crt Cathode Selector switch to the Chopped Blanking position and check that...

Страница 9: ...Set the oscilloscope Time Cm switch to 5 p Sec Cm 3 Adjust the oscilloscope triggering controls for a stable chopped waveform display Fig 2 2 4 Set the oscilloscope Crt Cathode Selector switch to the...

Страница 10: ...eck of an oscilloscope pre amplifier system use a 1 series or letter series plug in unit rather than the Type TU 7 When using the Type TU 7 to couple an input signal to the oscilloscope vertical ampli...

Страница 11: ...FUNCTION switch operate the dual trace switching circuit or the pulse generator circuit and will be discussed under those headings Detailed Description When TEST FUNCTION SW10 is set to the LOW LOAD p...

Страница 12: ...ching multivibrator V95A V95B and steering diodes V102A V102B Under normal conditions the circuit performs five general functions 1 When the TEST FUNCTION switch is set to ALTERNATE the switching mult...

Страница 13: ...Circuit Description Type TU 7 Fig 3 2 S im plified schematic o f Dual Trace Sw itching circuit show ing basic o p e ra tio n w ith TEST FUNCTION switch at ALTERNATE A 3 3...

Страница 14: ...output level applied to C22 drops from 0 1 volt to about ground set by R88 R89 Any signal now applied to the EXT INPUT connector is attenu ated by the 1000 1 attenuator at the input and applied throug...

Страница 15: ...Time Base B sweep When the TEST FUNCTION switch is set to the CHOPPED position the voltage in the grid circuit of the switching multivibrator sets the circuit for astable operation See Fig 3 5 The sw...

Страница 16: ...Circuit Description Type TU 7 Fig 3 5 Dual Trace Sw itching circuit show ing o pe ra tio n w ith TEST FUNCTION switch at CHOPPED...

Страница 17: ...uit When the TEST FUNCTION switch is set to the PULSE or PULSE position the rate generator operates as an astable circuit One complete multivibrator cycle turns the current switching transistors on an...

Страница 18: ...Circuit Description Type TU 7 Fig 3 7 S im plified schematic o f the Pulse G e n e rator circuit w ith TEST FUNCTION switch a t PULSE and AMPLITUDE control fu lly clock wise 3 8...

Страница 19: ...voltage on the crt display of the pulse With the AMPLITUDE control turned fully clockwise and the disconnect diodes conduct ing this output is approximately 67 2 volts at the junc tion of D75 and R76...

Страница 20: ...rted when Q55 turns on again The switching rate of the rate generator multivibrator is determined by the supply voltages and the resistor capacitor combinations in the base and emitter circuits The fr...

Страница 21: ...troubles not apparent during regular operation may be revealed and corrected during calibration CORRECTIVE MAINTENANCE General Information Removal or replacement procedures for most parts in the Type...

Страница 22: ...htly over the stud area of the strip Switch Replacement Individual wafers normally are not replaced in switch assemblies Replacement switiches may be ordered from Tektronix either unwired or with the...

Страница 23: ...k 1000 9 HIGH LOAD 2 4 k 1000 9 ALTERNATE CHOPPED 35 k 1000 10 LOW LOAD GAIN SET COMMON MODE ALTERNATE CHOPPED 12 1 k 1000 10 HIGH LOAD 1 5 k 1000 10 PULSE PULSE 4 4 k 12 k 1000 11 LOW LOAD GAIN SET C...

Страница 24: ...Connect the test instruments to the power line 4 Turn on all test instruments except the indicator oscillo scope 5 Set the indicator oscilloscope front panel controls as fol lows Horizontal Display T...

Страница 25: ...n adjustments and test points 4 Cheek VARIABLE Control o Reset the Type TU 7 controls to the positions given under Preliminary Instructions b Set the Time Base A triggering control to free run the swe...

Страница 26: ...ns f Check that the upper trace is locked to Time Base B 10 Check Chopped Operation a Remove the Calibrator signal b Set the TEST FUNCTION switch to CHOPPED c Set the indicator oscilloscope Horizontal...

Страница 27: ...ope power If the indicator oscilloscope has a plug in sensing switch in the plug in compartment pull forward on the switch plunger g Place the coupling capacitors item 9 under Equip ment Required on t...

Страница 28: ...counter sunk P pico or 10 1 2 dia diameter PHS pan head steel div division piv peak inverse voltage EMC electrolytic metal cased piste plastic EMT electroyltic metal tubular PMC paper metal cased ext...

Страница 29: ...Parts List Type TU 7 EXPLODED VIEW 6 2 CABLt HARNESS AND STRIP DETflIL...

Страница 30: ...u inch 210 840 1 LUG solder pot plain 5 333 791 1 PANEL front 6 366 220 1 KNOB AMPLITUDE charcoal Includes 213 020 1 SCREW set 6 32 x V8 inch HSS Mounting Hardware not included 210 590 1 NUT hex 3 8 3...

Страница 31: ...inch BHS 18 Resistor Mounting Hardware 210 478 1 NUT hex aluminum 5 u x 7 1 32 inch 210 601 1 EYELET brass tapered barrel 211 553 1 SCREW 6 32 x 1y2 inch RHS 211 507 1 SCREW 6 32 x 5 1 6 inch BHS 19...

Страница 32: ...LIP diode 30 136 0150 00 4 SOCKET 3 pin Mounting Hardware For Board not included 210 0586 00 6 NUT keps 4 40 x y 4 inch 211 0008 00 6 SCREW 4 40 x inch 31 179 0850 00 1 CABLE harness chassis 32 179 08...

Страница 33: ...5 281 518 47 pf Cer 500 v C47 283 051 0 0033 pf Cer 100 v 5 C48 283 004 0 02 pf Cer 150 v C49 283 026 0 2 pf Cer 25 v C50 283 026 0 2 pf Cer 25 v C52 281 525 470 pf Cer 500 v C55 281 518 47 pf Cer 500...

Страница 34: ...51 108 Replaceable by 2N2501 064 151 097 2N955 074 151 083 Selected from 2N964 Resistors Resistors are fixed composition 10 unless otherwise indicated R10 302 103 10k Y a w Rll 311 395 2 5 k Var VARIA...

Страница 35: ...Var AMPLITUDE R74 323 633 801 O A w Prec 1 R76 321 068 49 9 0 Vs w Prec 1 R78 321 289 10k y8 w Prec 1 R79 302 225 2 2 meg A w R80A 316 104 100 k A w R80B 316 104 100k A w R80C 316 104 100 k A w R80D...

Страница 36: ...25 k 10 w w w 5 R116 308 018 2 5 k 10 w w w 5 R121 324 603 27 8 k 1 w Prec 1 R122 308 025 20 k 10 w W W 5 R122 308 027 30 k 10 w W W 5 R123 312 642 5000 20 w w w 1 R124 301 623 62 k y2w 5 R125 303 433...

Страница 37: ...ns CONTROLS SETTINGS VARIABLE centered VERTICAL POSITION centered AMPLITUDE centered REPETITION RATE MED Input Signal none Pulse Gen ckt A ll other ckts TEST FUNCTION PULSE CHOPPED Test O scilloscope...

Страница 38: ......

Страница 39: ...CIRCUIT DIAGRAM rl Q T 0 Z i TLST UNIT...

Страница 40: ...t and component improvements to our instruments as soon as they are developed and tested Sometimes due to printing and shipping require ments we can t get these changes immediately into printed manual...

Страница 41: ...R21 521 0615 00 5 05 k 1 8 w Prec 1 2 R32 CHANGE 502 0222 00 TO 2 2 k 1 2 W 10 CVf 285 0627 00 0055 M E PTM 100 V CkS 285 0685 OO 0 2 2 pF REM 100 V CA9 285 0625 00 A7 pF REM 100 V R 5 5 0 1 0155 00...

Страница 42: ...TYPE 1M1 TENT S N 1 10 Page 2 of 2 SCHEMATIC CORRECTION PA R TIA L C IR C U IT D IA G R A M M 9 7 3 1 1 2 6 5...

Страница 43: ...Parts List Correction Change to R109 308 0093 00 12 K 8 W W W 5 M 9 7 3 3 H 6 5...

Страница 44: ...H W L 44 c0 74 t 6 0 6 7 d f w 4 rco 5 7 7 j 2 5 1 r v i 1y 6 w 4 V t r W r r 4 r i L V l t f f...

Отзывы: