![Tektronix Trimode TDP7700 Series Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/tektronix/trimode-tdp7700-series/trimode-tdp7700-series_technical-reference_1078231021.webp)
Figure 12: Operating voltage window (common mode, flex tip)
Voltage window with P77C292MM SMA adapter
The P77C292MM adapter operating voltage window is shown in the following figure.
Figure 13: Operating voltage window (P77C292MM SMA adapter)
The A operating voltage window range for the adapter of ±4.0 V is shown on the vertical axis and the B operating voltage
window range of ±4.0 V is shown on the horizontal axis. The dynamic range is also shown on the figure as the smaller
green square.
Similar to the operating voltage for the TekFlex solder-in tips, the dynamic range square labeled Small Signal AC can be
moved around within the limits of the operating voltage window by adjusting the A and B offset voltage settings. The A and
B offset voltage values determine the location of the center of the dynamic range square in the operating voltage window
plot. In C and D mode, only the common-mode and differential offsets are available. These operating windows are
illustrated by the figures below.
Figure 14: Operating voltage window (differential mode, P77C292MM adapter)
Figure 15: Operating voltage window (common mode, P77C292MM adapter)
Theory of operation
TDP7700 Series TriMode™ Probes Technical Reference
13