Complete the following steps to complete this procedure. Record the results on
the copy of the Calibration Data Sheet.
1.
Go to the Setup window of the logic analyzer and set the probe threshold
voltages to 2.5 V.
2.
Set the voltage source to 2.300 V.
3.
Start the logic analyzer and verify that it does not trigger.
4.
Increase the voltage in 10 mV steps, waiting at least 3 seconds between steps
to make sure that the logic analyzer continues to run without triggering.
Continue until the logic analyzer triggers and then record the voltage.
5.
Set the voltage source to 2.700 V.
6.
Start the logic analyzer and verify that it does not trigger.
7.
Decrease the voltage in 10 mV steps, waiting at least 3 seconds between
steps to make sure that the logic analyzer continues to run without triggering.
Continue until the logic analyzer triggers and then record the voltage.
8.
Add the two voltage values and divide by two. Verify that the result is
2.50 V ±75 mV. Record the voltage on the Calibration Data Sheet.
9.
Go to the Setup window and set the logic analyzer threshold voltages to –2.0 V.
10.
Repeat steps 3 through 8 for -2.200 V and -1.800 V.
11.
Add the two voltage values and divide by two. Verify that the result is
-2.00 V ±70 mV. Record the voltage on the Calibration Data Sheet.
12.
Repeat the procedure for each probe channel group that you want to verify.
20
TLA6400 Series Product Speci
fi
cations & Performance Veri
fi
cation