THM500 Series Performance Verification
Bench Test Instruments and Handheld Oscilloscopes Basic Service
11
Functional Test
The following steps verify basic instrument operation and reset the THM500
series instruments to factory default settings.
1.
Install fully charged batteries or use the THMCOM1 Communications
Adapter or the THM5AC Power Adapter optional accessories to provide
power.
2.
Press the
ON/OFF
button to turn the THM500 series instrument on.
3.
Press the
MENU ON/OFF
button to display the menus.
NOTE
. The THM575 can display the menus in more than one language. The
language selections are located in the Utility menu.
4.
Select
UTILITY
"
STATUS
"
DISPLAY
from the menus.
5.
Press the
SELECT
button. Verify that the correct THM500 series instrument
model number is displayed.
6.
Use the menus to select
RESET ALL
. It will appear underlined.
7.
Press the
SELECT
button and wait for the confirmation message; then press
the
SELECT
button again.
8.
Verify that the instrument returns to its power-on mode:
H
THM550, THM560, THM565, THM570, or THM570U instruments
return to METER mode.
H
THM571, THM575, or THM570U Option 3K instruments return to a
banner screen.