Getting Started
Figure 1-7: Removing the module
5.
If this empty slot will not be used, you must attach a blank panel to the
mainframe to control instrument cooling and EMI emissions.
a.
Align the blank panel with the open slot of the mainframe.
b.
Tighten the two screws to secure the blank panel to the mainframe.
A blank panel is available as an optional accessory for the
TG700 mainframe. Refer to
Accessories
. (See page 1-3.)
6.
If you want to install another module into the slot from which the module
is removed, perform
Installing a Module
starting at step 4. (See page 1-9,
Module Installation and Removal
Functional Checks
Use the following procedures to check that the TG700 modules installed in your
instrument are functional.
AG7 Module Functional
Check
Use the following procedure to check that the AG7 Audio Generator module
is functional.
Required Equipment.
Digital audio monitor
Oscilloscope
TG700 TV Signal Generator Platform User Manual
1–13