![Tektronix TDS7054 Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/tektronix/tds7054/tds7054_service-manual_1078193043.webp)
Specifications
TDS7104 & TDS7054 Service Manual
1- 21
Table 1- 11: Certifications and compliances
Category
Standards or description
EC Declaration of Conformity --
EMC
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European Union:
EN 61326 Emissions
1, 3
Class A Radiated and Conducted Emissions
EN 61326 Immunity
1
,
2
,4
IEC 1000-4-2
Electrostatic Discharge Immunity
±
4 kV contact discharge,
±
8 kV air discharge,
performance criterion B
IEC 1000-4-3
RF field immunity
3 V/m, 80 MHz to 1 GHz,
80% amplitude modulated with a 1 kHz sinewave
performance criterion A
IEC 1000-4-4
Electrical Fast Transient/Burst Immunity
1 kV on AC mains, 500 V on I/O cables, performance criterion B
IEC 1000-4-5
AC Surge Immunity
1 kV differential mode, 2 kV common mode,
performance criterion B
IEC 1000-4-6
RF Conducted Immunity
3 V, 150 kHz to 80 MHz, amplitude modulated with a
1 kHz sinewave, performance criterion A
IEC 1000-4-11
AC Mains Voltage Dips and Interruption Immunity
100% reduction for one cycle, performance criterion B
EN 61000-3-2
Power Harmonic Current Emissions
1
If interconnect cables are used, they must be low-EMI shielded cables such as the following
Tektronix part numbers or their equivalents: 012-0991-01, 012-0991-02 or 012-0991-03 GPIB
Cable; 012-1213-00 (or CA part number 0294-9) RS-232 Cable; 012-1214-00 Centronics
Cable; or LCOM part number CTL3VGAMM-5 VGA Cable.
2
The performance criteria for when the instrument is subjected to the conditions described
above are defined as follows:
A —
≤
0.2 division waveform displacement or
≤
0.4 division increase in peak-to-peak noise
B — temporary, self-recoverable degradation or loss of performance is allowed, but no change
of actual operating state or loss of stored data is allowed
C — temporary loss of function is allowed provided that the function is self recoverable or can
be restored by the operation of the controls
3
Radiated emissions may exceed the levels specified in EN 61326 when this instrument is
connected to a test object.
4
USB mouse and keyboard only, performance criterion C. Normal USB keyboard or mouse
operation can be restored by unplugging and reconnecting the USB connector to the
instrument.
FCC
Radiated and conducted emissions do not exceed the levels specified in FCC47 CFR, Part 15,
Subpart B, for Class A equipment.
Содержание TDS7054
Страница 4: ......
Страница 14: ...Table of Contents x TDS7104 TDS7054 Service Manual...
Страница 18: ...Service Safety Summary xiv TDS7104 TDS7054 Service Manual...
Страница 22: ...Preface xviii TDS7104 TDS7054 Service Manual...
Страница 56: ...Operating Information 2 12 TDS7104 TDS7054 Service Manual...
Страница 60: ...Theory of Operation 3 4 TDS7104 TDS7054 Service Manual...
Страница 64: ...Performance Verification 4 4 TDS7104 TDS7054 Service Manual...
Страница 76: ...Brief Procedures 4 16 TDS7104 TDS7054 Service Manual...
Страница 86: ...Performance Tests Semi Automated Method 4 26 TDS7104 TDS7054 Service Manual...
Страница 144: ...Performance Tests 4 84 TDS7104 TDS7054 Service Manual...
Страница 160: ...Maintenance 6 6 TDS7104 TDS7054 Service Manual...
Страница 165: ...Removal and Installation Procedures TDS7104 TDS7054 Service Manual 6 11 Figure 6 2 Knob removal...
Страница 230: ...Troubleshooting 6 76 TDS7104 TDS7054 Service Manual...
Страница 232: ...Repackaging Instructions 6 78 TDS7104 TDS7054 Service Manual...
Страница 238: ...Electrical Parts List 8 2 TDS7104 TDS7054 Service Manual...