Tektronix TDS2012 Скачать руководство пользователя страница 5

tds2012_oscilloscope_guide.doc 

5/18

 

Channel menus 

The yellow CH1 MENU and blue CH2 MENU buttons are used to cause the oscilloscope to 
either display or hide the signal from the corresponding channel (yellow trace for CH1 and blue 
trace for CH2).   In addition, when the particular channel is selected, a softkey menu for that 
channel will be activated.  The particular softkey options are: 

a)

 

The <Coupling> softkey allows the user to select from 

DC

 (displays/measures both AC and 

DC signal components),

 AC 

(displays/measures only AC signal components), or 

Ground

 

(displays flat line at 0 V) options.  

b)

 

The <BW Limit> softkey allows the user to toggle between 

Off

 (full 100 MHz) and 

On

 

(limit signals to 20 MHz).  We’ll always leave this 

Off

c)

 

The <Volts/Div> softkey allows the user to toggle between 

Coarse

 (use oscilloscope 

defaults, e.g., 10 V/div, 5 V/div, 2 V/div, 1 V/div …) and 

Fine

 (allows for V/div selections 

between the defaults). 

d)

 

The <Probe> softkey allows the user to select from 

1X

10X

100X

, and 

1000X

 to 

compensate for attenuation in the probes.  Use 

1X

 (no attenuation) when the P2200 probe 

switch is set to 

1X

 or a BNC-BNC cable is connected to the oscilloscope.  Use 

10X

 when 

the P2200 probe switch is set to 

10X

.  The other settings are meant for different probes. 

e)

 

The <Invert> softkey allows the user to toggle between 

Off

 (display signal as-is) and 

On

 

(inverts signal).  We’ll usually leave this 

Off

Volts/Div knobs 

These knobs, directly below yellow CH1 MENU and blue CH2 MENU buttons, are used to 
adjust the vertical scales of the CH1 and CH2 traces.  The settings for each channel will be 
shown at the lower left corner of the display.  These numbers refer to the number of volts 
between each horizontal gridline on the display 

for that channel’s signal

.  The separate knobs 

allow you to compare two signals at different levels of magnification. 

Sec/Div knob 

Use this knob to adjust the horizontal or time scale of the display (same for both channels).  The 
current setting will be shown at the bottom center of the display.  This number refer to the 
number of seconds between each vertical gridline on the.  Adjusting this knob allows you to vary 
the size of the time slice visible on the display.  When the Sec/Div control is set to 100 ms/div or 
slower and the trigger mode is set to Auto, the oscilloscope enters the scan acquisition mode. In 
this mode, the waveform display updates from left to right.  There is no trigger or horizontal 
position control of waveforms during scan mode. 

Position knobs 

There are three position knobs available on the oscilloscope.  Under the “VERTICAL” menu, the 

POSITION knobs (one for each channel) allows the user to move the signal for that channel up 

or down on the display, without changing the voltage scale.  Essentially, this knob moves where 
the ground (0 V) reference level for each channel is placed.  The reference level is indicated on 
the screen, e.g., a little yellow arrow labeled “1” is shown to the left of screen for CH1.  Under 
the “HORIZONTAL” menu, the POSITION knob allows users to move all displayed signals left 
and right on the display, without changing the time scale.  This knob moves where the time 
reference point is placed (little white arrow labeled on top of screen).    The oscilloscope displays 
information regarding this position in the upper right corner. 

Dr. Montoya 

Rev. 3/16/09 

StockCheck.com

Содержание TDS2012

Страница 1: ...cilloscope_guide doc 1 18 Tektronix TDS 2012 Digital Storage Oscilloscope User s Guide Authors Contributors Thomas Montoya James Free Jake Gunderson John Preheim Dr Montoya Rev 3 16 09 S t o c k C h e...

Страница 2: ...5 Sec Div knob 5 Position knobs 5 Set to zero 6 Help 6 Run Stop 6 Single Sequnce 6 Horizontal menu 6 Math Menu 7 Menus 10 Save Recall 8 Measure 9 Acquire 10 Utility 11 Cursor 13 Display 15 Trigger Con...

Страница 3: ...UN STOP PROBE COMP 5V 1kHz VERTICAL HORIZONTAL TRIGGER ON OFF SWITCH 0 1 Figure 1 Front panel layout of TDS2012 Digital Storage Oscilloscope In order to make measurements the oscilloscope must be conn...

Страница 4: ...laced little white arrow labeled on top of screen The oscilloscope displays information regarding this position in the upper right corner Printing oscilloscope display 1 Press the UTILITY button in ME...

Страница 5: ...will be shown at the lower left corner of the display These numbers refer to the number of volts between each horizontal gridline on the display for that channel s signal The separate knobs allow you...

Страница 6: ...ed with angle brackets such as Autoset These are links to other topics Turn the Help Scroll Position knob to move the highlight from one link to another Push the Show Topic softkey button to display t...

Страница 7: ...oosing Window expands the Window Zone to cover the entire screen When you change between the Main Window Zone and Window views the oscilloscope erases any waveform saved on the screen Math Menu Pushin...

Страница 8: ...the waveform in the time domain use Autoset or the control knobs to center the waveform vertically Also make sure that the entire waveform is visible on the display Horizontally the FFT will only cons...

Страница 9: ...the oscilloscope acquires a time domain waveform containing frequency components that are greater than the Nyquist frequency The frequency components that are above the Nyquist frequency are undersam...

Страница 10: ...controls and the cursors to magnify and measure the FFT spectrum Once you have properly dealt with any aliasing issues you are free to use the knobs to position and zoom the display and use the cursor...

Страница 11: ...toggle between choices of CH1 and CH2 Type which allows the user to cycle through the eleven types of measurements shown in Table 6 and Back which returns you to the top MEASURE softkey menu Pressing...

Страница 12: ...pe_guide doc 12 18 Table 6 Types of measurements 3 Push the ACQUIRE button to bring up a softkey menu with the selections listed in Table 7 Table 7 ACQUIRE softkey menu Dr Montoya Rev 3 16 09 S t o c...

Страница 13: ...button to bring up a softkey menu with the selections listed in the first column of Table 8 Some these selections bring up subchoices as listed in the second column of Table 8 Table 8 UTILITY softkey...

Страница 14: ...now redefined as cursor control knobs The voltage Delta between the cursors Cursor 1 position and Cursor 2 position are shown on the screen under the softkey selections see bottom part of Table 9 When...

Страница 15: ...ions listed in Table 10 Table 10 DISPLAY softkey menu Trigger Controls You can define the trigger see Table 11 by pushing the TRIG MENU button and front panel soft keys While in each trigger type the...

Страница 16: ...tds2012_oscilloscope_guide doc 16 18 Dr Montoya Rev 3 16 09 S t o c k C h e c k c o m...

Страница 17: ...tds2012_oscilloscope_guide doc 17 18 Dr Montoya Rev 3 16 09 S t o c k C h e c k c o m...

Страница 18: ...igger signal FORCE TRIG Completes an acquisition regardless of an adequate trigger signal This button has no effect if the acquisition is already stopped TRIG VIEW Displays the trigger waveform in pla...

Отзывы: