Calibration
6.
When the Probe Setup screen appears,
select
Clear ProbeCal
, and then select
Calibrate Probe
.
The probe calibration routine begins.
When the routine completes, the Probe
Status changes from Running to Pass.
Close the noti
fi
er and begin using your
probe.
If the Probe Status changes to Fail:
Check that the adapters are securely
attached to the probe tips and to the
PROBE COMP connectors on the
oscilloscope.
Verify that the selected channel on
the left side of the Probe Setup menu
matches the channel that the probe is
plugged into.