Reference Notes
Figure 12: Measuring differential current and nulls
3.
Measure the current. A waveform above the baseline indicates the conductor
with the conventional current
fl
ow (I
1
, in the direction of the probe arrow),
is carrying the greater current. Conventional current
fl
ows from positive to
negative.
4.
To adjust for a current null, adjust the current in one of the conductors until
the displayed measurement is zero.
TCPA300/400 Ampli
fi
ers and TCP300A/400 Series Current Probes User Manual
25