Speci
fi
cations
Speci
fi
cations
The speci
fi
cations in the tables in this section are valid under the following conditions:
The probe has been calibrated at an ambient temperature of 23 °C ±5 °C.
The probe is connected to a host instrument with an input impedance of 50
Ω
.
The probe must have a warm-up period of at least 20 minutes and be in an environment that does not exceed the limits
described. (See Table 1.)
Speci
fi
cations for the TCP202A current probe fall into three categories: warranted, typical, and nominal characteristics.
Warranted Characteristics
Warranted characteristics describe guaranteed performance within tolerance limits or certain type-tested requirements. Warranted
characteristics have checks in the
Performance Veri
fi
cation
section.
Table 1: Warranted electrical characteristics
Characteristic
Description
DC gain accuracy
<3% (typical <1% at +23 °C, ±5 °C)
Rise time (10% to 90%)
≤
7.0 ns
Bandwidth
DC to 50 MHz
TCP202A Current Probe Instruction Manual
39
Содержание TCP202A
Страница 1: ...x TCP202A 50 MHz 15 A AC DC Current Probe ZZZ Instruction Manual P071300300 071 3003 00...
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv TCP202A Current Probe Instruction Manual...
Страница 62: ...Specifications Figure 2 Typical input impedance versus frequency 42 TCP202A Current Probe Instruction Manual...
Страница 63: ...Specifications Figure 3 Maximum peak pulse versus pulse width TCP202A Current Probe Instruction Manual 43...
Страница 65: ...Specifications Figure 4 Probe dimensions TCP202A Current Probe Instruction Manual 45...