Operation
6.
Edit the contents of the table to match
the loss characteristics of the equipment
you are correcting for. Click
OK
when
you are
fi
nished.
NOTE.
You can enter either positive or
negative values into the table. But in contrast
to the External Gain value setting, positive
values represent a loss and negative values
represent a gain.
For detailed information on using the Table
Editor, select
Help
>
User Manual
and look
in the Index for
External Loss Tables
.
7.
You can change the title for each external
loss table. Select the title box and type
in a new title.
Note that the title is only a label. It is
not tied to the
fi
le name of any table you
have loaded or saved.
8.
Specify whether table corrections are
applied to
all
acquired data or only to
traces in the Spectrum, Spectrogram,
Spurious and Amplitude vs Time
displays.
54
RSA6100A Series Quick Start User Manual