Tektronix P80318 Скачать руководство пользователя страница 2

creasing pressure beyond this point acceler-
ates probe tip wear and is not recommended.

Optimum probe tip

pressure

Probe tip travel

Pr

obe

tip

pressure

H

When the probe engages the second spring,
the probe sends a control signal to the 80A02
EOS/ESD protection module to enable the
signal path.

H

Do not scrape or otherwise drag the probe
tips on the board traces or DUT leads in an
attempt to improve signal contact. Make
sure that the contact area is clean before
using the probe.

H

Do not bend or shape the ground wire.
Improper bending of the wire can reduce
signal acquisition bandwidth or accuracy.

H

Cover the probe tip when not in use with the
vinyl probe cover.

H

When using the P80318 probe with auto-
mated probing stations, do not exceed
0.040 inches travel from initial DUT contact.

Removing/Installing the Ground Wire

Use knife tip to lift

end of wire

Step 1

Set probe tips to maximum gap

Pull out old

wire with
tweezers

Step

2

Use tweezers to hold the new

wire. Insert the wire end ~1/3

length into the small hole

Step 3

1

2

Step 4

Apply light pressure

against the wire to

keep in position

Pivot wire to position it on top of the support block.

Then use flat blade end of adjustment tool to push

wire down between block and probe tip.

Step 5

Push end of wire down

into hole. Insert screw
after wire is positioned

Ground wire is under tension during
installation and can easily spring away

Clamping the Probe for Fixtured Probing

Clamp

zone

Specifications

Table 1: Electrical

Characteristic

Specification

Probe impedance

50

Ω

±

1.5

Ω

, single ended,

each polarity.

Probe delay

1 ns typical from the female
SMA connector end to the
probe tip.

Attenuation

1X

Risetime, probe head
and cable

19.4 ps or better, one-way
transit, differential drive, with
minimum probe tip spacing (for
a 100 ps window with 50%
threshold crossing at center of
screen).

Bandwidth, probe head
and cable, calculated

18 GHz or better, using above
risetime calculation and
assuming 0.35 product of
risetime and bandwidth.

Maximum voltage input

±

5.0 V (DC + AC

p--p

)

RF input connector

Precision 26 GHz SMA female
connector

Control signal connector
type

Audio mono phone mini-jack

Table 2: Environmental mechanics

Characteristic

Specification

Probe dimensions

Length

127.0 mm (5.0 in)

Height

33.0 mm (1.30 in)

Width

28.0 mm (1.10 in)

Weight (probe only)

0.075 kg (0.165 lbs)

Altitude

Operating

3,000 m (9,843 ft)

Non-operating

12,190 m (40,000 ft)

Ambient temperature

Operating

--15

_

C to +55

_

C

(+5

_

F to +131

_

F)

Non-operating

--62

_

C to +85

_

C

(--80

_

F to +185

_

F)

Humidity

Operating

20% to 80% relative humidity
with a maximum wet bulb
temperature of +29

_

C at or

below +40

_

C,

non-condensing. (Upper limit
derates to 45% relative
humidity at +40

_

C.)

Отзывы: