Table 2: Electrical
Characteristic
Specification
Probe impedance
50
Ω
+
±
2%
Attenuation
1X
RF input connector
Precision 26 GHz SMA
female connector
Control signal connector
type
Audio mono phone mini-jack
Analog bandwidth
Probe head
<3 dB insertion loss, DC to
20 GHz
Cable
DC to 20 GHz
Probe fall time
(measured with 80E04 or
SD24 TDR modules)
< 38 ps
Typical reflected TDR fall
time < 28 ps
Probe delay
1 ns typical from the
connector end to the probe
tip
This is a stable impedance
section which can be
calibrated and used as a
reference point.
Maximum voltage input
±
5.0 V
Module activation probe
pressure
200 to 260 grams
Accessories
The P8018 probe includes the following items:
S
Special 1 meter SMA cable
with a parallel control line)
174-4879-xx
S
20 ground clips
016-1936-xx
WARRANTY SUMMARY
Tektronix warrants that the products that it
manufactures and sells will be free from defects in
materials and workmanship for a period of one (1)
year from the date of shipment. If a product
proves defective within the respective period,
Tektronix will provide repair or replacement as
described in the complete warranty statement.
To arrange for service or obtain a copy of the
complete warranty statement, please contact your
nearest Tektronix sales and service office.
EXCEPT AS PROVIDED IN THIS SUMMARY
OR THE APPLICABLE WARRANTY
STATEMENT, TEKTRONIX MAKES NO
WARRANTY OF ANY KIND, EXPRESS OR
IMPLIED, INCLUDING WITHOUT
LIMITATION THE IMPLIED WARRANTIES
OF MERCHANTABILITY AND FITNESS FOR
A PARTICULAR PURPOSE. IN NO EVENT
SHALL TEKTRONIX BE LIABLE FOR
INDIRECT, SPECIAL OR CONSEQUENTIAL
DAMAGES.