Performance Veri
fi
cation
Figure 28: Reverse the power supply polarity on the probe inputs
8.
Record the output voltage (on the second DMM) as V
out
2.
9.
Calculate the gain as follows: (V
out
1 - V
out
2) ÷ (V
in
1 - V
in
2).
10.
Verify that the gain is 0.4, ±2.0%.
11.
Record the calculated gain for the 2.5X setting on the test record.
Gain Check at 12.5X Attenuation
12.
Set the attenuation on the probe to 12.5X.
13.
Repeat steps 2 through 9, but in step 3, set each power supply to +0.7 V (+1.4 V differential total).
14.
Verify that the gain is 0.08, ±2.0%.
15.
Record the calculated gain on the test record.