Performance Veri
fi
cation
Output Offset Zero
By terminating the two probe SMA inputs with 50
Ω
, this procedure tests the zero output voltage of the probe. The probe
output is measured at the SMA connector on the front of the 80A03 interface.
1.
Connect the equipment as shown. (See Figure 26.)
2.
Connect two 50
Ω
terminations to the two probe SMA inputs on the probe, and plug the probe into the 80A03 module.
3.
Connect the cable from the DMM to the SMA connector that is located below the 80A03 channel that you plugged
the probe into.
Figure 26: Setup for the output offset zero test
4.
Set the Vterm source to Ext on the probe. Leave the external termination control voltage inputs open. This sets the
termination voltage to zero.
5.
Set the multimeter to read DC volts.
6.
Verify that the output voltage is 0 V, ±2.5 mV for both the 2.5X and 12.5X attenuation settings.
7.
Record the results on the test record.