Low-inductance Grounding
Placing a ground plane on top of a package being probed can
minimize ground lead length and inductance. See Figure 4.
Figure 4: Low-inductance Grounding
Attach a thin piece of copper on top of the package and connect it to
the package ground connection. Use the low-inductance ground lead
provided with the probe to keep the ground lead length as short as
possible.
This method is very useful when making many measurements on the
same package. Using a ground plane on the package makes probing
the package easier, and avoids adding unnecessary ground lead
length.