Display Waveform or Trace Data
The
fi
gure to the right shows the frequency
domain window’s trace indicator.
1.
An
RF
trace indicator is placed at the
Reference Level.
2.
A capital
M
appears if the maximum
trace is turned on.
3.
A capital
A
appears if the average trace
is turned on.
4.
A capital
N
appears if the normal trace
is turned on.
5.
The small
m
appears if the minimum
trace is turned on.
Orange highlighting indicates the currently
selected trace. In the
fi
gure to the right,
the small
m
, which stands for the minimum
trace, is highlighted. This indicates that the
minimum trace is currently selected.
Detection Types
The MDO3000 calculates FFTs with a 1,000 to ~2,000,000 point output, depending on the acquisition settings. It then
reduces that FFT output into a 750 pixel-wide display. This means that approximately 1 to 2,000 FFT points get compressed
into each pixel column. The MDO3000 gives you several choices as to how this compression is done. The choices are:
+peak, sample, average, and -peak. The
fi
gure below illustrates how these detection methods work in a 5:1 compression,
where
fi
ve points are reduced to each pixel column.
1.
FFT points
2.
Decimation
3.
+Peak: Uses the highest amplitude point
in each interval.
4.
Sample: Uses the
fi
rst point in each
interval.
MDO3000 Series Oscilloscopes User Manual
119
Содержание MDO3012
Страница 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Страница 2: ......
Страница 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Страница 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Страница 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...