Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-96
2470-901-01 Rev. A /
May
2019
Example
*RST
SOUR:FUNC VOLT
SOUR:VOLT:RANG 20
SENS:FUNC "CURR"
SENS:CURR:RANG 100e-6
SOUR:SWE:VOLT:LOG .1, 10, 20, 1e-3, 1, FIXED
INIT
Reset the instrument to its defaults.
Set the source function to voltage.
Set the source range to 20 V.
Set the measure function to current.
Set the current range to 100 μA.
Set up a log sweep that sweeps from 0.1 to 10 V in 20 steps with a source delay of 1 ms, a sweep count of 1,
and a fixed source range.
Start the sweep.
Also see
STATus subsystem
The STATus subsystem controls the status registers of the instrument. For additional information on
the status model, see
(on page B-1).
:STATus:CLEar
This function clears event registers.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:STATus:CLEar
Details
This command clears the event registers of the Questionable Event and Operation Event Register
set. It does not affect the Questionable Event Enable or Operation Event Enable registers.
Example
:STATus:CLEar
Clear the bits in the registers.
Also see
(on page A-2)