Models 707B and 708B Switching Matrix Reference Manual
Section 7: TSP command reference
707B-901-01 Rev. B / January 2015
7-59
Details
This function resets the following attributes to factory default settings:
•
digio.trigger[
N
].mode
•
digio.trigger[
N
].pulsewidth
•
digio.trigger[
N
].stimulus
It also clears
digio.trigger[
N
].overrun
.
Example
digio.trigger[3].mode = 2
digio.trigger[3].pulsewidth = 50e-6
digio.trigger[3].stimulus = digio.trigger[5].EVENT_ID
print(digio.trigger[3].mode, digio.trigger[3].pulsewidth,
digio.trigger[3].stimulus)
digio.trigger[3].reset()
print(digio.trigger[3].mode, digio.trigger[3].pulsewidth,
digio.trigger[3].stimulus)
Set the digital I/O trigger line 3 for a falling edge with a pulsewidth of 50 microseconds.
Use digital I/O line 5 to trigger the event on line 3.
Reset the line back to factory default values.
Output before reset:
2.000 5.0000000e-05 5.000
Output after reset:
0.000 1.0000000e-05 0.000
Also see
(on page 7-55)
(on page 7-57)
(on page 7-59)
digio.trigger[N].stimulus
This attribute selects the event that causes a trigger to be asserted on the digital output line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Digital I/O trigger
N
reset
Save setup
Create configuration script
0
Usage
triggerStimulus
= digio.trigger[
N
].stimulus
digio.trigger[
N
].stimulus =
triggerStimulus
triggerStimulus
The event identifier for the triggering event.
N